FDA Medical Device Recalls (openFDA)
Cook Inc.: Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
- Recall number
- Z-0975-2020
- Recalling firm
- Cook Inc.
- Product
- Sof-Flex Double Pigtail Ureteral Stent Set, Global Product No. G15076 G14906 G14840 G14865 G17128
- Status
- Terminated
- Initiated
- 2019-01-04
- Posted
- not on file
- Terminated
- 2020-05-29
- Reason
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity
- 13388 total
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Lot, serial or model codes
- NS7949093 NS8077953 NS8077954 NS8077955 8070691 8183342 8183343 8070693 8172134 8187189 NS8070692 8424304 7990841
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FAD
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28