Medtronic Inc: Medtronic Surgical Membrane Oxygenator. Mis-printed Label: Medtronic 1-4500-2A, Recommended maximum blood flow range 5.0 - 8.5 L/min, Sterilized Using Ethylene Oxide, REF 61399402655, Manufacturer: Medtronic Inc., Minneapolis, MN 55432-5604. Surgical Membrane Oxygenators are intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures. Additionally, the Integral Membrane Oxygenators are intended to cool and warm the blood by means of the integral heat exchange.

Recall number
Z-0975-2009
Recalling firm
Medtronic Inc
Product
Medtronic Surgical Membrane Oxygenator. Mis-printed Label: Medtronic 1-4500-2A, Recommended maximum blood flow range 5.0 - 8.5 L/min, Sterilized Using Ethylene Oxide, REF 61399402655, Manufacturer: Medtronic Inc., Minneapolis, MN 55432-5604. Surgical Membrane Oxygenators are intended to be used in an extracorporeal perfusion circuit to oxygenate and remove carbon dioxide from the blood during routine cardiopulmonary bypass procedures. Additionally, the Integral Membrane Oxygenators are intended to cool and warm the blood by means of the integral heat exchange.
Status
Terminated
Initiated
2008-10-17
Posted
2009-02-09
Terminated
2011-04-19
Reason
Medtronic has identified an issue with the labeling of the I-4500 Silicone Oxygenators (P/N 61399402655). Lot numbers of this oxygenator manufactured between August 2007 and September 2008 have an incorrect "Recommended maximum blood flow range" on the label. The label on the device states a range of 5.0 - 8.5 L/min but should state a range of 5.0 - 6.5 L/min. These oxygenators are acceptable t
Root cause
Error in labeling
Action
A consignee letter " Urgent Medical Deive Notice" dated October 17, 2008 was sent to the impacted customers on October 17, 2008. The letter described the issue and product and confirmed the correct maximum internal flow of 6.5 L/min for the I-4500 Silicone Oxygenators. The letter states the product will not be removed from the field. Customers were instructed to complete the" Medtronic Product Notification Certificate (FCA 0903)" acknowledging receipt of the information and fax the completed certificate to 763-391-9603. Please direct questions about the recall to your Medtronic Sales Representative or call 763-391-9124.
Quantity
408
Distribution
Worldwide Distribution USA including states of CA, CO, FL, IL, IN, KY, LA, MD, MI, MN, MO, NV, NJ, NY, OH, OK PA, TN, TX, UT, VA, and WI, and countries of LATVIA, ITALY, ARMENIA, RUSSIAN FEDERATION, POLAND, CHILE and CHINA.
Lot, serial or model codes
Lot Numbers: 10058759, 9984675, 9956368, 9916066, 989741, 9875810, 9866275, 9854094, 9828585, 9812351, 9421573, 9169873, 9044915, 8956269, 8938159, 8885755 and 8765243.   Serial Numbers: IH31608 - IH32031.
510(k) numbers
not on file
PMA numbers
not on file
Product code
DTZ
Firm FEI
3001504994
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register