Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: Siemens Primeview 3i medical charged-particle radiation therapy system, Part No: 7341428, with software version 3i. Siemens AG, Concord, CA.

Recall number
Z-0974-2008
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens Primeview 3i medical charged-particle radiation therapy system, Part No: 7341428, with software version 3i. Siemens AG, Concord, CA.
Status
Terminated
Initiated
2007-10-02
Posted
2008-04-30
Terminated
2008-12-24
Reason
Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
Root cause
Software design
Action
On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
Quantity
6 units
Distribution
Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US.
Lot, serial or model codes
Serial Numbers: 102301, 104201, 00E0815462E6, 5602, 48701, 57401
510(k) numbers
["K060226"]
PMA numbers
not on file
Product code
IYE
Firm FEI
2937457
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register