FDA Medical Device Recalls (openFDA)
Medtronic Navigation, Inc.: Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
- Recall number
- Z-0973-2014
- Recalling firm
- Medtronic Navigation, Inc.
- Product
- Medtronic O-arm Imaging System Mobile X-ray system. Model No. BI-700-00027 and BI-700-00028 The 0-arm Imaging System is a mobile x-ray system designed for 2D fluoroscopic and 3D imaging and is intended to be used where a physician benefits from 2D and 3D information of anatomic structures and objects with high x-ray attenuation such as bony anatomy and metallic objects. The 0-ann Imaging System is compatible with certain Image Guided Surgery Systems.
- Status
- Terminated
- Initiated
- 2013-07-25
- Posted
- 2014-02-28
- Terminated
- 2019-04-17
- Reason
- Review of on-system electronic product labeling and Instructions for comply with applicable Use in response to a customer inquiry uncovered gaps in compliance FDA standards with the requirements for information to be provided to users in 21CFR 1020.30(h)
- Root cause
- Radiation Control for Health and Safety Act
- Action
- Medtronic Navigation Inc. issued Notiifcation on 3/4/14 and included thee errata sheet to all affected customers updating the missing and incorrect information and ensure that labeling is now in compliance with 21 CFR Part 1020.30(h). This will be followed by a software update that contains the same information on the errata sheet, which serves to correct the electronic version of the Instructions for Use. The paper errata sheet will be sent via traceable means and the software upgrade will be performed by Medtronic Service either through remote access or by site visit of Medtronic service personnel. For further questions please call (978) 698-6000.
- Quantity
- 449 in US
- Distribution
- USA (nationwide) Distribution
- Lot, serial or model codes
- Model No. BI-700-00027 and BI-700-00028 Serial number range: 00102 through 00753 and B13080721, B13080722, B13090729, B13120763, B13120764, B13120766, B13120768, B13120769
- 510(k) numbers
- ["K092564"]
- PMA numbers
- not on file
- Product code
- OXO
- Firm FEI
- 3004785967
- Firm city
- Littleton
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28