Skip to the content

FDA Medical Device Recalls (openFDA)

Keystone Dental Inc: PrimaConnex Tapered Implant TC, SD Internal Connection 3.5 x 10mm Catalog Number: l5613K. Implants in support for fixed bridgework.

Recall number
Z-0973-2013
Recalling firm
Keystone Dental Inc
Product
PrimaConnex Tapered Implant TC, SD Internal Connection 3.5 x 10mm Catalog Number: l5613K. Implants in support for fixed bridgework.
Status
Terminated
Initiated
2013-03-11
Posted
2013-03-20
Terminated
2013-09-26
Reason
Mislabeled: Cover screw that was in the package was a WD cover screw and not a SD cover screw as labeled
Root cause
Packaging process control
Action
Keystone Dental initially called customers on 3/11/13 and forms were completed to document these calls. In addition, a letter sent via Federal Express/or hand delivered to all distributors and customers. Product requested to be returned.
Quantity
120 units
Distribution
Worldwide Distribution-USA including DC and the states of CA, FL, IN, MA, MN, NJ, NY, NH, PA, WI, and NY, and the countries of Republic of China, Germany, and France.
Lot, serial or model codes
Lot Number: 12857
510(k) numbers
["K051614"]
PMA numbers
not on file
Product code
NDP
Firm FEI
3005990499
Firm city
Burlington
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register