Skip to the content

FDA Medical Device Recalls (openFDA)

THOR Photomedicine Ltd: NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3) Product Code S2190 - XL

Recall number
Z-0972-2024
Recalling firm
THOR Photomedicine Ltd
Product
NovoTHOR Whole Body Light Pod, Gen 3.0 (1) Product Code S2188 - Reg (2) Product Code S2189 - Reg PRX (3) Product Code S2190 - XL
Status
Open, Classified
Initiated
2023-11-20
Posted
2024-02-01
Terminated
not on file
Reason
One of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or close the canopy
Root cause
Component design/selection
Action
On November 20, 2023, the firm notified customers via IMPORTANT MEDICAL DEVICE CORRECTION letters sent through email. A service agent representing THOR will contact customers within the next eight weeks to schedule the fitting of replacement ball studs at no cost to the customer. Customers who continue to use the NovoTHOR should be aware that it may potentially cause harm should the failure occur while users/patients are lifting or lowering the canopy. Customers should warn all users of potential risk, ensure someone is present to lift and lower the canopy, ensure the canopy is lifted and lowered gently and slowly, and ensure someone is present in the room or immediately outside the room during use.
Quantity
61 (41 US; 20 OUS)
Distribution
Domestic distribution to the states of Arizona California Colorado Connecticut Florida Georgia Hawaii Idaho Indiana Maryland New Jersey New York Oregon Pennsylvania Texas Washington Wisconsin. Foreign distribution to Australia, Canada, Italy, Malta, Norway, Switzerland, UAE, and UK.
Lot, serial or model codes
(1) UDI-DI 05060494130503 Serial Numbers 4004 4009 4013 4017 4019 4021 4024 4026 4028 4029 4035 4037 4042 4043 4044 4046 4047 4048 4054 4065 4066 4072 4073 4074 4076 4077 4078 4079 4080 4081 4082 4083 4085 4086 4087 4090 4091 4096 (2) UDI-DI 05060494130527 Serial Number 4068 (3) UDI-DI 05060494130510 Serial Numbers 4002 4003 4006 4010 4012 4027 4030 4040 4038 4023 4045 4051 4052 4060 4057 4064 4056 4069 4063 4070 4089 4014
510(k) numbers
not on file
PMA numbers
not on file
Product code
ILY
Firm FEI
3003899624
Firm city
Amersham
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register