Skip to the content

FDA Medical Device Recalls (openFDA)

Medimop Medical Projects Ltd.: Vial2BagDC, Product Number 6070112. The 20mm Vial2Bag DC device, is intended for use in healthcare facilities or in-home environment by the patient or care-giver to aid and support prescribed treatment and therapy. The device consists of the Vial2Bag piercing spike and cover, the IV Port connector and an integrated Vial Adapter (20mm) for access to the drug/solution vial.

Recall number
Z-0972-2018
Recalling firm
Medimop Medical Projects Ltd.
Product
Vial2BagDC, Product Number 6070112. The 20mm Vial2Bag DC device, is intended for use in healthcare facilities or in-home environment by the patient or care-giver to aid and support prescribed treatment and therapy. The device consists of the Vial2Bag piercing spike and cover, the IV Port connector and an integrated Vial Adapter (20mm) for access to the drug/solution vial.
Status
Terminated
Initiated
2017-10-06
Posted
not on file
Terminated
2024-01-23
Reason
Medimop is recalling certain lots of this product Vial2Bag DC due to the potential of the IV Port separating from the Via12Bag DC body during the drug reconstitution I IV Set attachment process.
Root cause
Process control
Action
The recalling firm's sent an Urgent Medical Device Recall Letter dated October 13, 2017. was supplied to the consignee which contains the instructions for the recall process to the consignee. The notification letter will be sent to the Distributor via email. The notification letter instructs the distributor to notify the Customer/Users via the Medimop consignee recall letter. For further questions, please call (610) 594-4353.
Quantity
453,600 units
Distribution
US Distribution
Lot, serial or model codes
Lots: 9507,9508,9509,9510,9511,9512,9513.
510(k) numbers
["K140730","K170095"]
PMA numbers
not on file
Product code
LHI
Firm FEI
3000223297
Firm city
Raanana
Firm state
not on file
Firm country
Israel

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register