Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Healthcare Diagnostics: Siemens CLINITEST hCG -Pregnancy Test (qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG) test system. Package of 25. Product Code 1760/ Catalog# 06484105). For in vitro diagnostic use as a qualitative method in the rapid detection of human chorionic gonadotropin (hCG) in urine specimens.

Recall number
Z-0972-2012
Recalling firm
Siemens Healthcare Diagnostics
Product
Siemens CLINITEST hCG -Pregnancy Test (qualitative), Radio immunoassay,Human Chorionic Gonadotropin (hCG) test system. Package of 25. Product Code 1760/ Catalog# 06484105). For in vitro diagnostic use as a qualitative method in the rapid detection of human chorionic gonadotropin (hCG) in urine specimens.
Status
Terminated
Initiated
2011-12-15
Posted
2012-02-07
Terminated
2015-03-30
Reason
A potential for false positive or increase in borderline results with patients and commercially available negative control material.
Root cause
Process control
Action
Siemens issued Overnight Mailings of Urgent: Recall Notices on December 15, 2011 for delivery beginning December 16, 2011 to Distributors and end users. Customer bulletin has been issued instructing customers to discontinue use of product and discard impacted inventory. Customers are to contact their distributor regarding credit for unused product. For technical questions contact Siemens Technical Solutions Center at 877-229-3711.
Quantity
48,578 Kits (25 tests / kit)
Distribution
Worldwide Distribution -- US, Austria, Germany, Great Britain, Denmark, Norway, Sweden, Poland, Netherlands, Luxembourg, Belgium, Finland, France, Greece, Ireland, Italy, Portugal, Qatar, Saudi Arabia, Slovakia, Slovenia, and Switzerland.
Lot, serial or model codes
Lot Expiration 028619, 6/8/2012; 028924, 7/5/2012; 029997, 10/7/2012; 030240, 11/6/2012; 030431, 11/24/2012; 030602, 12/8/2012; 030826, 12/24/2012; 030998, 1/11/2013; 031332, 2/4/2013; 031460, 2/19/2013; 031662, 3/6/2013; 031798, 3/20/2013; 031924, 3/30/2013; 032180, 4/28/2013; and 032106, 4/19/2013.
510(k) numbers
["K032563"]
PMA numbers
not on file
Product code
JHI
Firm FEI
2432235
Firm city
Tarrytown
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register