Skip to the content

FDA Medical Device Recalls (openFDA)

Madison Polymeric Engineering: Draco Enzymatic Deep-Cleaning Pads, Product Number HY0305Z Product Usage: Multi-tiered enzymatic detergent contains lipase, carbohydrase, amylase and protease. It effectively dissolves blood, fat, tissue, protein and most other forms of organic material Ready to use. Non-toxic, latex-free and neutral pH. Safe to use on all surgical instruments, flexible endoscopes and probes.

Recall number
Z-0969-2019
Recalling firm
Madison Polymeric Engineering
Product
Draco Enzymatic Deep-Cleaning Pads, Product Number HY0305Z Product Usage: Multi-tiered enzymatic detergent contains lipase, carbohydrase, amylase and protease. It effectively dissolves blood, fat, tissue, protein and most other forms of organic material Ready to use. Non-toxic, latex-free and neutral pH. Safe to use on all surgical instruments, flexible endoscopes and probes.
Status
Terminated
Initiated
2018-12-14
Posted
not on file
Terminated
2019-12-31
Reason
The product is contaminated with Pseudomonas.
Root cause
Nonconforming Material/Component
Action
Cygnus sent an Urgent Medical Device Recall notification letters dated December 14, 2018 to customers. The letter identified the affected product, problem and actions to be taken. For questions contact Customer Service at 800-990-7489.
Quantity
4800 pads
Distribution
Worldwide Distribution - US states: CA, FL, IL, IN, NJ, NY, PA, and TN. The products were distributed to the following foreign countries: Canada
Lot, serial or model codes
Lot numbers: EX20190419P EX20190420P EX20190423P EX20190424P
510(k) numbers
not on file
PMA numbers
not on file
Product code
MNL
Firm FEI
3001451466
Firm city
Branford
Firm state
CT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register