Skip to the content

FDA Medical Device Recalls (openFDA)

Becton Dickinson & Company: BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.

Recall number
Z-0968-2018
Recalling firm
Becton Dickinson & Company
Product
BD Vacutainer K2 EDTA (K2E) 3.6mg Blood Collection Tube (13 x 75 mm x 2.0 mL). Catalog Number: 367841. Used for whole blood hematology determinations, immunohematology testing and blood donor screening.
Status
Terminated
Initiated
2017-08-29
Posted
not on file
Terminated
2019-05-16
Reason
A limited portion of the lot was manufactured with less than the required amount of K2EDTA additive, an anticoagulant, which may cause erroneous results that could lead to specimen recollection and potential delay of treatment or misdiagnosis/mismanagement of treatment and serious complications, such as failure to detect thrombocytosis or postponing surgery.
Root cause
Under Investigation by firm
Action
In August of 2017 All US direct consignees were contacted via e-mail and/or UPS mail. An Urgent Product Removal Recall letter was issued, identifying the affected device and the reason for the recall. It also requested customers to review their inventory and quarantine any affected devices, which will be returned to the firm for replacement. The Customer Recall Response Form should be completed and returned regardless of whether there is any affected devices on hand. For questions or concerns please contact Customer Service at 1-855-215-4992.
Quantity
8,494,000
Distribution
Nationwide Distribution
Lot, serial or model codes
Catalog No. 367841 Lot No. 6279849 UDI: (01)30382903678410
510(k) numbers
["BK050036"]
PMA numbers
not on file
Product code
JKA
Firm FEI
2243072
Firm city
Franklin Lakes
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register