FDA Medical Device Recalls (openFDA)
Ortho-Clinical Diagnostics: VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate container used in making automatic (auto)-dilutions on the VITROS 250/350 Chemistry Systems. The VITROS 250 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Microslides. The VITROS 350 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides.
- Recall number
- Z-0968-2014
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- VITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage: Mixing Cup Array is a disposable, intermediate container used in making automatic (auto)-dilutions on the VITROS 250/350 Chemistry Systems. The VITROS 250 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products Microslides. The VITROS 350 Chemistry System is intended for use in the in vitro quantitative, semi quantitative, and qualitative measurement of a variety of analytes of clinical interest, using VITROS Chemistry Products MicroSlides.
- Status
- Terminated
- Initiated
- 2013-10-10
- Posted
- 2014-02-10
- Terminated
- 2018-05-18
- Reason
- Ortho Clinical Diagnostics issued a product correction notification for VITROS Chemistry Mixing Cup Arrays utilized on VITROS 250 and VITROS 350 Chemistry Systems. Increase in Software error due to affected Mixing Cup Arrays.
- Root cause
- Nonconforming Material/Component
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification dated October 10, 2013 to all affected customer. The letter identifed the affected product, problem and actions to be taken. Customers were instructed to complete and return the attached Confirmation of Receipt form. For questions contact the Customer Technical Services at 1-800-421-3311.
- Quantity
- 6380 units (each unit contains 48 arrays with 10 microcups)
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
- Lot, serial or model codes
- 1) VITROS 250 (Product Code 8132086; Model # 3332); 2) VITROS 350 (Product Code 6802153; Model # 3330); 3) Mixing Cup Arrays (Product Code 1631779; Model # n/a)
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- JRB
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28