Del Mar Reynolds Medical, Ltd.: Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.

Recall number
Z-0968-2013
Recalling firm
Del Mar Reynolds Medical, Ltd.
Product
Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
Status
Terminated
Initiated
2013-01-15
Posted
2013-03-20
Terminated
2013-05-14
Reason
A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.
Root cause
Process control
Action
Spacelabs sent an URGENT-MEDICAL DEVICE CORRECTION letter, dated 16 January 2013, to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers until a Spacelabs representative replaces the affected product: 1) Immediately cease the use of BleaseSirius and BleaseFocus Anesthesia Workstations that shipped after 31 October 2012. 2) Immediately cease use of Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation containing a CAS I/II Absorber from such Service Kit. 3) Advise your staff of this situation Customers can call Spacelabs Healthcare at 1-800-522-7025 and select 2 for Technical support.
Quantity
108 total (17 units in the US and 91 units outside the US)
Distribution
Worldwide Distribution - USA Nationwide including the states of New Mexico, North Carolina and the countries of Argentina, China, Mexico, and Taiwan.
Lot, serial or model codes
serial numbers:  ABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482, ABSO-004483, ABSO-004487, ABSO-004488, ABSO-004489, ABSO-004490, ABSO-004491, ABSO-004492, ABSO-004493, ABSO-004494, ABSO-004495, ABSO-004497, ABSO-004499, ABSO-004500, ABSO-004503, ABSO-004504, ABSO-004505, ABSO-004506, ABSO-004507, ABSO-004508, ABSO-004510, ABSO-004511, ABSO-004512, ABSO-004513, ABSO-004526, ABSO-004528, ABSO-004529, ABSO-004530, ABSO-004531, ABSO-004532, ABSO-004533, ABSO-004534, ABSO-004535, ABSO-004536, ABSO-004537, ABSO-004538, ABSO-004539, ABSO-004540, ABSO-004541, ABSO-004542, ABSO-004543, ABSO-004544, ABSO-004545, ABSO-004564, ABSO-004566, ABSO-004567, ABSO-004568, ABSO-004569, ABSO-005883, ABSO-005884, ABSO-005885, ABSO-005886, ABSO-005887, ABSO-005945, ABSO-005947, ABSO-005948, ABSO-005949, ABSO-005950, ABSO-100023, ABSO-100024, ABSO-100038, ABSO-100039, ABSO-100040, ABSO-100041, ABSO-100042, ABSO-100043, ABSO-100070, ABSO-100071, ABSO-100072, ABSO-100073, ABSO-100074, ABSO-100075, ABSO-100076, ABSO-100077, ABSO-100078, ABSO-100079, ABSO-100080, ABSO-100081, ABSO-100082, ABSO-100083, ABSO-100084, ABSO-100085, ABSO-100086, and ABSO-100087.   Firm added ABSO 004581 - which was inSpacelaabs inventory, in transit to a Field Service Engineer, at the time the recall was initiated. the nonconforming absorber has been replaced with a conforming part and this absober serial number was added to the US Spacelabs Recall tracking.  Five incorrect serialnumbers were included in the original serial numbers provided to FDA. This was a result of a data entry error. An updated US and Rest of the World customer/Serial number was provided to FDA 3/25/2013. The correction to the Rest of the World report is as follows: Incorrect SN Reported to FDA Correct, Nonconforming SN ABSO 005883 ABSO-004514 ABSO 005884 ABSO-004515 ABSO 005885 ABSO-004516 ABSO 005886 ABSO-004517 ABSO 005887 ABSO-004518
510(k) numbers
["K003251","K101850"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3002807992
Firm city
Hertford
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register