Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Sternal Saw II Sterile Blades for the Sarns Sternal Saw System, Catalog number 5589, Terumo Cardiovascular System Corp, Ann Arbor, MI. The sternal saw blades are indicated for use in medial sternotomoties. The saw blades are used wit the Sarns Sternal Saw System.

Recall number
Z-0968-2011
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Sternal Saw II Sterile Blades for the Sarns Sternal Saw System, Catalog number 5589, Terumo Cardiovascular System Corp, Ann Arbor, MI. The sternal saw blades are indicated for use in medial sternotomoties. The saw blades are used wit the Sarns Sternal Saw System.
Status
Terminated
Initiated
2010-10-20
Posted
2011-01-26
Terminated
2011-12-14
Reason
The firm discovered that some outer box labels contain incorrect information regarding the method of sterilization. The label incorrectly indicates the product was ethylene oxide sterlized but the product actually was sterilzed with gamma radiation.
Root cause
Labeling design
Action
Terumo Cardiovascular Systems Corporation firm sent out an 'URGENT MEDICAL DEVICE CORRECTION: SAFETY ADVISORY" letter dated October 20, 2010, to the Director of Cardiac Operating Room Services or Operating Room Nurse Manager.. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review the Safety Advisory, assure that all users were aware of the notice, and confirm receipt of the Safety Advisory by faxing the attached Customer Response Form to the fax number on the form. For questions regarding this recall call 1-800-521-2818.
Quantity
42,650 units, 4265 boxes of 10 each
Distribution
US, Indonesia, UAE, Hong Kong, Japan, Singapore, Malaysia, Taiwan, Belgium, Uraguay, Venezuela, Columbia, Mexico, Guam, Vietnam and Thailand.
Lot, serial or model codes
0505258, 0510335, 0519809, 0524042, 0531107, 0533191, 0536444, 0538524, 0541634, 0545325, 0547867, 0552336, 0561753, 0561751, 0561752, 0557559, 0564337, 0570411, 0571613, 0573009, 0577882, 0581075, 0584965, 0587430, 0585731, 0585730, 0585732, 0598360, 0601970 and 0605113.
510(k) numbers
["K935391"]
PMA numbers
not on file
Product code
GFA
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register