ORTHO SOLUTIONS, LTD.: SYSTEM 26. Part Numbers: OS900035-NS; OS900036-NS; OS900037-NS; OS900038-NS; OS900040-NS; OS900041-NS; OS900042-NS; OS900043-NS; OS900047-NS; OS900048-NS; OS900045-NS; OS900046-NS; OS900050-NS; OS900051-NS; OS900044-NS; OS900049-NS; OS900053-NS; OS900054-NS; OS900055-NS; OS900056-NS; OS900058-NS; OS900059-NS; OS900057-NS.

Recall number
Z-0967-2018
Recalling firm
ORTHO SOLUTIONS, LTD.
Product
SYSTEM 26. Part Numbers: OS900035-NS; OS900036-NS; OS900037-NS; OS900038-NS; OS900040-NS; OS900041-NS; OS900042-NS; OS900043-NS; OS900047-NS; OS900048-NS; OS900045-NS; OS900046-NS; OS900050-NS; OS900051-NS; OS900044-NS; OS900049-NS; OS900053-NS; OS900054-NS; OS900055-NS; OS900056-NS; OS900058-NS; OS900059-NS; OS900057-NS.
Status
Terminated
Initiated
2017-10-10
Posted
not on file
Terminated
2018-08-31
Reason
Following the identification of corrosion residue found inside the cannulated devices in Forefoot and Midfoot trays Orhto Solutions Group has taken the decision to recall ALL x106 System 26 Forefoot and Midfoot Implants and Instruments tray from the US market.
Root cause
Process control
Action
Ortho Solutions notified customers of the recall via letter 10/10/2017. The recall notification letter was accompanied by instructions to Distributors and a Product Removal Form.
Quantity
120
Distribution
U.S.
Lot, serial or model codes
Lot Numbers: 1088402;1088403;1088404;1088405;1088406;1091608;1091646; 1091648;1091628;1088393;1091658; 1088394;1091586; 1088395;1091598;1088396;1088397;1091623; 1088398;1091618; 1088399;1091596; 1091599; 1088401;1091597; 1088337;1091622;1093232;1088985;1091542; 1088986;1091401;1088987;1091581;1091543; 1088989;1091589;1088990;1091614; 1088991;1091402;1091562
510(k) numbers
["K163489"]
PMA numbers
not on file
Product code
HWC
Firm FEI
3008951116
Firm city
Maldon
Firm state
not on file
Firm country
United Kingdom

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register