FDA Medical Device Recalls (openFDA)
Amendia, Inc: Ceres-C Standalone Cervical Interbody The product is a stand-alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at the one level from C2-T1.
- Recall number
- Z-0966-2017
- Recalling firm
- Amendia, Inc
- Product
- Ceres-C Standalone Cervical Interbody The product is a stand-alone anterior cervical interbody fusion device indicated for use in skeletally mature patients with degenerative disc disease (DDD) with accompanying radicular symptoms at the one level from C2-T1.
- Status
- Terminated
- Initiated
- 2016-11-22
- Posted
- 2017-01-09
- Terminated
- 2017-02-28
- Reason
- Absent tantalum market pin.
- Root cause
- Process control
- Action
- The firm, Amendia, contacted the consignee and issued a Return Material Authorization for the affected product on 8/18/2016. Please contact Regulatory and Quality Consultant for Amendia at 404-610-7215 or by email to bbrosseau@securebme.com with any questions or concerns regarding this recall notification.
- Quantity
- 6 devices
- Distribution
- US Distribution to GA only.
- Lot, serial or model codes
- Part Number 08-170-1010 - Lot Number 134926; Part Number 08-170-1011 - Lot Number 134927; Part Number 08-170-1006 - Lot Number 134929; Part Number 08-170-1008 - Lot Number 134931
- 510(k) numbers
- ["K152972"]
- PMA numbers
- not on file
- Product code
- OVE
- Firm FEI
- 3007970248
- Firm city
- Marietta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28