Given Imaging Ltd.: Bravo pH Monitoring capsule, UPC 7290101361695 (FGS# 0312) and UPC 7290101361688 (FGS# 0313). Product Usage: The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children form 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry.

Recall number
Z-0966-2012
Recalling firm
Given Imaging Ltd.
Product
Bravo pH Monitoring capsule, UPC 7290101361695 (FGS# 0312) and UPC 7290101361688 (FGS# 0313). Product Usage: The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children form 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry.
Status
Terminated
Initiated
2012-01-11
Posted
2012-02-06
Terminated
2012-09-14
Reason
Failure of the Bravo capsule to attach to the esophagus or alternatively, failure of the capsule to detach from the placement device.
Root cause
Nonconforming Material/Component
Action
Given Imaging sent a Urgent Product Recall letter to all affected consignees. The letter identified the product, description of problem, actions to be taken and advised customers to read this important information packet in its entirety. Customers were advised to complete the Customer Response Sheet included and return per the instructions provided. Also attached to the letter was a Customer Bulletin: Update to Vacuum Settings. For questions call 1-800-448-3644 option 3.
Quantity
21,885 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Austria, Belgium, Brazil, Denmark, Canada, Cyprus, Finland, Germany, Hong Kong, India, Indonesia, Israel, Italy, Japan, Mexico, Netherlands, Norway, New Zealand, Panama, Portugal, Singapore, Sri Lanka, South Africa, Spain Thailand, Turkey, United Kingdom and Venezuela.
Lot, serial or model codes
FGS-0312 (5 pack), Lot numbers: 11775Q, 12064Q, 12227Q, 12333Q, 12389Q, 12440Q, 12475Q, 12494Q, 12643Q, 12545Q, 12769Q, 12770Q, 12840Q, 12896Q, 12897Q, 12928Q, 12989Q, 13026Q, 13056Q, 13109Q, 13109Q, 13162Q, 13164Q, 13208Q, 13233Q, 13246Q, 13282Q, 13283Q, 13371Q, 13390Q, 13462Q, 13461Q, 13540Q, 13569Q, 13634Q, 13679Q, 13695Q, 13700Q, 13712Q, 13716Q, 13753Q, 13777Q, 13786Q, 13798Q, 13799Q, 13813Q, 13836Q, 13867Q, 13894Q, 13937Q, 13979Q, 14064Q, 14101Q, 14120Q, 14141Q, 14230Q, 14255Q, 14265Q, 14284Q, 14338Q, 14400Q, 14419Q, 14423Q, 14458Q, 14491Q, 14542Q, 1457QQ, 14572Q, 14570Q, 14611Q, 14634Q, 14734Q, 14770Q, 14805Q, 14825Q, 14826Q, 14871Q, 15030Q, 15051Q, 15071Q, 15078Q, 15114Q, 15158Q, 15221Q, 15324Q, 15406Q, 15401Q, 15500Q, 15603Q, 15630Q, 15672Q, 15679Q, 15711Q, 15787Q, 15860Q, 15912Q, 15938Q, 15971Q, 15998Q, 16115Q, 16144Q, 16192Q, 16221Q, 16320Q, 16290Q, 16354Q, 16382Q, 16395Q, 16449Q, 16503Q, 16511Q, 16580Q, 16606Q, 16607Q, 16673Q, 16762Q, 16867Q, 16762Q & 16847Q and FGS-0313 (single pack), Lot numbers: 12647Q, 12765Q, 12797Q, 12902Q, 13234Q, 13458Q, 13608Q, 13632Q, 13635Q, 13751Q, 13868Q, 13907Q, 14011Q, 14065Q, 14231Q, 14339Q, 14402Q, 14612Q, 14761Q, 14783Q, 14865Q, 14984Q, 15222Q, 15538Q, 15698Q, 15766Q, 15937Q, 16392Q, 16194Q, 16398Q, 16707Q, 16879Q, 17058Q & 16912Q. Recall expanded on 02/20/2012 to include additional lot numbers: 17102Q, 17188Q and 17227Q.
510(k) numbers
["K102543"]
PMA numbers
not on file
Product code
LLT
Firm FEI
3003141635
Firm city
Norcross
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register