FDA Medical Device Recalls (openFDA)
Given Imaging Ltd.: Bravo pH Monitoring capsule, UPC 7290101361695 (FGS# 0312) and UPC 7290101361688 (FGS# 0313). Product Usage: The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children form 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry.
- Recall number
- Z-0966-2012
- Recalling firm
- Given Imaging Ltd.
- Product
- Bravo pH Monitoring capsule, UPC 7290101361695 (FGS# 0312) and UPC 7290101361688 (FGS# 0313). Product Usage: The Bravo pH Monitoring System is intended to be used for gastroesophageal pH measurement and monitoring of gastric reflux in adults and children form 4 years of age. The Bravo pH Capsule can be attached following either endoscopy or manometry.
- Status
- Terminated
- Initiated
- 2012-01-11
- Posted
- 2012-02-06
- Terminated
- 2012-09-14
- Reason
- Failure of the Bravo capsule to attach to the esophagus or alternatively, failure of the capsule to detach from the placement device.
- Root cause
- Nonconforming Material/Component
- Action
- Given Imaging sent a Urgent Product Recall letter to all affected consignees. The letter identified the product, description of problem, actions to be taken and advised customers to read this important information packet in its entirety. Customers were advised to complete the Customer Response Sheet included and return per the instructions provided. Also attached to the letter was a Customer Bulletin: Update to Vacuum Settings. For questions call 1-800-448-3644 option 3.
- Quantity
- 21,885 units
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Austria, Belgium, Brazil, Denmark, Canada, Cyprus, Finland, Germany, Hong Kong, India, Indonesia, Israel, Italy, Japan, Mexico, Netherlands, Norway, New Zealand, Panama, Portugal, Singapore, Sri Lanka, South Africa, Spain Thailand, Turkey, United Kingdom and Venezuela.
- Lot, serial or model codes
- FGS-0312 (5 pack), Lot numbers: 11775Q, 12064Q, 12227Q, 12333Q, 12389Q, 12440Q, 12475Q, 12494Q, 12643Q, 12545Q, 12769Q, 12770Q, 12840Q, 12896Q, 12897Q, 12928Q, 12989Q, 13026Q, 13056Q, 13109Q, 13109Q, 13162Q, 13164Q, 13208Q, 13233Q, 13246Q, 13282Q, 13283Q, 13371Q, 13390Q, 13462Q, 13461Q, 13540Q, 13569Q, 13634Q, 13679Q, 13695Q, 13700Q, 13712Q, 13716Q, 13753Q, 13777Q, 13786Q, 13798Q, 13799Q, 13813Q, 13836Q, 13867Q, 13894Q, 13937Q, 13979Q, 14064Q, 14101Q, 14120Q, 14141Q, 14230Q, 14255Q, 14265Q, 14284Q, 14338Q, 14400Q, 14419Q, 14423Q, 14458Q, 14491Q, 14542Q, 1457QQ, 14572Q, 14570Q, 14611Q, 14634Q, 14734Q, 14770Q, 14805Q, 14825Q, 14826Q, 14871Q, 15030Q, 15051Q, 15071Q, 15078Q, 15114Q, 15158Q, 15221Q, 15324Q, 15406Q, 15401Q, 15500Q, 15603Q, 15630Q, 15672Q, 15679Q, 15711Q, 15787Q, 15860Q, 15912Q, 15938Q, 15971Q, 15998Q, 16115Q, 16144Q, 16192Q, 16221Q, 16320Q, 16290Q, 16354Q, 16382Q, 16395Q, 16449Q, 16503Q, 16511Q, 16580Q, 16606Q, 16607Q, 16673Q, 16762Q, 16867Q, 16762Q & 16847Q and FGS-0313 (single pack), Lot numbers: 12647Q, 12765Q, 12797Q, 12902Q, 13234Q, 13458Q, 13608Q, 13632Q, 13635Q, 13751Q, 13868Q, 13907Q, 14011Q, 14065Q, 14231Q, 14339Q, 14402Q, 14612Q, 14761Q, 14783Q, 14865Q, 14984Q, 15222Q, 15538Q, 15698Q, 15766Q, 15937Q, 16392Q, 16194Q, 16398Q, 16707Q, 16879Q, 17058Q & 16912Q. Recall expanded on 02/20/2012 to include additional lot numbers: 17102Q, 17188Q and 17227Q.
- 510(k) numbers
- ["K102543"]
- PMA numbers
- not on file
- Product code
- LLT
- Firm FEI
- 3003141635
- Firm city
- Norcross
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28