Skip to the content

FDA Medical Device Recalls (openFDA)

Biocare Medical Llc: MUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA; Product Usage: For In Vitro Diagnostic Use. This antibody may be used as a tool for the identification and the sub classification of lymphoid malignancies : Multiple myeloma oncogene-1 (MUM-1) is a protein encoded by MUM-1 gene. MUM-1 protein is expressed in the nuclei and cytoplasm of plasma cells and a small percentage of germinal center (GC) B-cells located in the light zone. MUM-1 [BC5] labels MUM-1protein in centrocytes and their progeny, plasma cells, activated T-cells and a wide spectrum of hematolymphoid neoplasms derived from these cells. MUM-1 has been reported to play an important role in mediating B-cell activation and differentiation.

Recall number
Z-0965-2015
Recalling firm
Biocare Medical Llc
Product
MUM-1 (Multiple myeloma oncogene-1), catalog number PRM352 AA; Product Usage: For In Vitro Diagnostic Use. This antibody may be used as a tool for the identification and the sub classification of lymphoid malignancies : Multiple myeloma oncogene-1 (MUM-1) is a protein encoded by MUM-1 gene. MUM-1 protein is expressed in the nuclei and cytoplasm of plasma cells and a small percentage of germinal center (GC) B-cells located in the light zone. MUM-1 [BC5] labels MUM-1protein in centrocytes and their progeny, plasma cells, activated T-cells and a wide spectrum of hematolymphoid neoplasms derived from these cells. MUM-1 has been reported to play an important role in mediating B-cell activation and differentiation.
Status
Terminated
Initiated
2014-12-12
Posted
2015-01-13
Terminated
2015-03-30
Reason
Three lots of MUM-1 may not meet their labeled shelf life, as a drop in staining intensity over time has been observed.
Root cause
Incorrect or no expiration date
Action
Biocare sent a Medical Device Recall notifications letter to all customers by traceable method. The letters identified the affected product, problem and actions to be taken. Users are asked to immediately discontinue used of the affected lots and dispose of them according to applicable national, state and local laws. Users are requested to complete and return the response form so that proper credit can be issues 800-799-9499.
Quantity
59
Distribution
Worldwide Distrinution - US Nationwide in the states of TX, WI, MO, WV, SC, MA, WA, TN, AZ, CA, IL, NY, LA and countries of Kuwait, India, Venezuela, New Zealand, Turkey, Uruguay.
Lot, serial or model codes
Lot 061013, Exp 2016/06; Lot 073013, Exp. 2016/07; Lot 071614, Exp. 2017/07.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NJT
Firm FEI
3004140393
Firm city
Concord
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register