FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories, Inc.: D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.
- Recall number
- Z-0965-2014
- Recalling firm
- Bio-Rad Laboratories, Inc.
- Product
- D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.
- Status
- Terminated
- Initiated
- 2013-11-22
- Posted
- 2014-02-10
- Terminated
- 2014-03-20
- Reason
- D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.
- Root cause
- Software design
- Action
- BIO-RAD sent an Urgent Medical Device Correction letter dated November 22, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to continue to use the affected update kit disk(s) included with the reorder pack(s), but manually track the number of injections on the cartridge, reviewing QC results closely throughout the life of the reorder pack or destroy the affected disk(s) included with these reorder pack(s) and replace with the new Update Kit disk(s) The new Update Kit disks will elminate the problem and may be used with all reorder packs. The replacement disks do not contain reagent or calibrator information; after installing teh new Update Kit disk, they ust manually update the component information using instructions provided with the disk. For questions regarding this recall call 510-741-6265.
- Quantity
- 2906 kits (318 Kits in US; 2588 kits internationally)
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.
- Lot, serial or model codes
- Catalog number: 220-0201: Reorder Pack Lot No./Expiration date: 20030322 3/31/2014 20030337 3/31/2014 64000081 5/31/2014 64000331 5/31/2014 64000332 5/31/2014 64000888 08/31/2014 64001344 04/30/2014 64002239 11/30/2014 ( 64002240 08/3112014 64002241 11/30/2014
- 510(k) numbers
- ["K031043"]
- PMA numbers
- not on file
- Product code
- LCP
- Firm FEI
- 2915274
- Firm city
- Hercules
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28