Skip to the content

FDA Medical Device Recalls (openFDA)

Cook Inc.: PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10

Recall number
Z-0962-2020
Recalling firm
Cook Inc.
Product
PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
Status
Terminated
Initiated
2019-01-04
Posted
not on file
Terminated
2020-05-29
Reason
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Root cause
Process control
Action
On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
Quantity
3929
Distribution
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Lot, serial or model codes
7832498 7832499 7832500 7925385 7925387 7930517 8090868 8287793 8287794 8287799 8400121 8548813 8548815 NS8548814 7760235 7881599 8062433 8074027 8290383 8290384 8404559 8404562 7999663 8077582 8077583 8065959 8073949 8074014 8074015
510(k) numbers
["K082536"]
PMA numbers
not on file
Product code
OCY
Firm FEI
1820334
Firm city
Bloomington
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register