FDA Medical Device Recalls (openFDA)
Cook Inc.: PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
- Recall number
- Z-0962-2020
- Recalling firm
- Cook Inc.
- Product
- PTFE Wire Guide, Guidewire, Global Product Number 638813 635413-10
- Status
- Terminated
- Initiated
- 2019-01-04
- Posted
- not on file
- Terminated
- 2020-05-29
- Reason
- The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
- Root cause
- Process control
- Action
- On January 4, 2019, the firm distributed Urgent Medical Device Recall letters to customers via courier mail. Customers were informed that affected lots may contain a wire guide that was incorrectly loaded into the wire guide holder.
- Quantity
- 3929
- Distribution
- Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
- Lot, serial or model codes
- 7832498 7832499 7832500 7925385 7925387 7930517 8090868 8287793 8287794 8287799 8400121 8548813 8548815 NS8548814 7760235 7881599 8062433 8074027 8290383 8290384 8404559 8404562 7999663 8077582 8077583 8065959 8073949 8074014 8074015
- 510(k) numbers
- ["K082536"]
- PMA numbers
- not on file
- Product code
- OCY
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28