Skip to the content

FDA Medical Device Recalls (openFDA)

MedAmicus,Inc: Medamicus FlowGuard 16 FR Valved Peelable Introducer Kit, Catalog Number 10532-011

Recall number
Z-0962-03
Recalling firm
MedAmicus,Inc
Product
Medamicus FlowGuard 16 FR Valved Peelable Introducer Kit, Catalog Number 10532-011
Status
Terminated
Initiated
2003-06-06
Posted
2003-07-02
Terminated
2004-07-02
Reason
Stability testing, after accelerated aging of the device, showed that the handle of device did not meet strength specifications.
Root cause
Other
Action
The consignees were sent recall letters, dated June 6, 2003 or June 20, 2003, that requested the return of the products.
Quantity
20 units
Distribution
The products were distributed to 4 consignees in Georgia, Missouri, New York and North Carolina. Most of the product was shipped to a distributor in New York which probably shipped to other states. There was no direct distribution to foreign or U.S Government accounts.
Lot, serial or model codes
Lot Number W23152
510(k) numbers
["K021044"]
PMA numbers
not on file
Product code
DYB
Firm FEI
2183787
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register