FDA Medical Device Recalls (openFDA)
Cook Inc.: Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent Reference Part Number (RPN): 003700 GPN: G14837
- Recall number
- Z-0960-2021
- Recalling firm
- Cook Inc.
- Product
- Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal Shock Wave Lithotripsy (E.S.W.L.) and leaving an indwelling ureteral stent Reference Part Number (RPN): 003700 GPN: G14837
- Status
- Terminated
- Initiated
- 2021-01-04
- Posted
- not on file
- Terminated
- 2021-09-16
- Reason
- Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
- Root cause
- Nonconforming Material/Component
- Action
- Cook issued Urgent Medical Device Recall Letter on Jan 4 , 2021 via email/mail. Letter states reason for recall, health risk and action to take: If affected products are currently indwelling in a patient, consider scheduling a follow-up with the patient to remove and/or replace the product, based on the status and preferences of each individual patient. 2.Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all distribution and use of this product. 3.Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. Note: Unaffected products that are returned will not be credited. 4.Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter. Even if you do not have affected product(s) on hand, you must still complete the Acknowledgement and Receipt Form and return via fax (812.339.7316) or email(FieldActionsNA@CookMedical.com). 5.This notice must be shared with appropriate personnel, including down to the user level, with in your organization or with any organization where the potentially affected devices have been transferred. 6.Immediately report adverse events to Cook Medical Customer Relations by phone at800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (EasternTime), or by email to: CustomerRelationsNA@CookMedical.com.
- Quantity
- 42 units
- Distribution
- FL, KY, MD, NS, NY, PA, TX Foreign: Canada, Switzerland
- Lot, serial or model codes
- Lot Numbers: 13024536 13089895 NS13089897 UDI: (01)00827002148376(17)230303(10)13024536 (01)00827002148376(17)230327(10)13089895 (01)00827002148376(17)230327(10)NS13089897
- 510(k) numbers
- ["K162109"]
- PMA numbers
- not on file
- Product code
- FAD
- Firm FEI
- 1820334
- Firm city
- Bloomington
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28