Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: Symbia T and Symbia T2 computer tomography systems. These systems are used to perform SPECT, CT, and SPECT + CT scans and imaging studies.

Recall number
Z-0960-2015
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Symbia T and Symbia T2 computer tomography systems. These systems are used to perform SPECT, CT, and SPECT + CT scans and imaging studies.
Status
Terminated
Initiated
2014-03-05
Posted
2015-01-13
Terminated
2016-12-23
Reason
There are confirmed reports of a component coming loose inside the CT enclosure and damaging the system. components.
Root cause
Component change control
Action
The firm, Siemens, sent an undated "Customer Safety Advisory Notice" letter to all customers. The described the product, problem and actions to be taken. The letters inlcuded instructions for customers to: 1) continue to use their systems while waiting for the field correction to be scheduled and performed; 2) ensure that the safety advisory is placed in the system's instruction for use; and, 3) closely observe the patient during studies. The letter also instructs customers of record to forward a copy of the letter to the new owner if the equipment was sold or transferred. Customers with any questions about this issue can contact Siemens Molecular Imaging at 800-888-7436 (US).
Quantity
295 systems
Distribution
Worldwide Distribution: US Nationwide including Puerto Rico; Austria, Australia, Bangladesh, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Czech Republic, Ecuador, Egypt, Finland, France, Germany, Great Britain, Hungary, Ireland, India, Iran, Italy, Japan, Mexico, Netherlands, New Zealand, Norway, Philippines, Poland, Portugal, Russia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, and Tunisia.
Lot, serial or model codes
1) Symbia T; Part Number: 10275007; Serial Numbers: 1055 , 1057 , 1058 , 1060 , 1061 , 1062 , 1063 , 1064 , 1065 , 1066 , 1067 , 1068 , 1069 , 1070 , 1071 , 1072 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1080 , 1081 , 1082 , 1083 , 1084 , 1085 , 1086 , 1087 , 1088 , 1089 , 1090 , 1091 , 1092 , 1094 , 1095 , 1096 , 1097 , 1098 , 1099 , 1100 , 1101 , 1102 , 1103 , 1104 , 1105 , 1106 , 1107 , 1108 , 1109 , 1110 , 1111 , 1112 , 1113 , 1114 , 1115 , 1116 , 1117 , 1118 , 1119 , 1121 , 1122 , 1123 , 1124 , 1125 , 1126 , 1127 , 1128 , 1129 , 1130 , 1131 , 1132 , 1133 , 1134 , 1135 , 1136 , 1137 , 1138 , 1139 , 1140 , 1141 , 1142 , 1143 , 1144 , 1145 , 1146 , 1147 , 1148 , 1149 , 1150 , 1151 , 1152 , 1153 , 1154 , 1155 , 1156 , 1157 , 1158 , 1159 , 1160 , 1161 , 1162 , 1163 , 1164 , 1165 , 1166 , 1167 , 1168 , 1169 , 1170 , 1171 , 1172 , 1173 , 1175 , 1176 , 1177 , 1178 , 1179 , 1180 , 1181 , 1182 , 1183 , 1184 , 1185 , 1186 , 1187 , 1188 , 1189 , 1190 , 1191 , 1193 , 1194 , 1195 , 1196 , 1197 , 1198 , 1199 , 1200 , 1201 , 1202 , 1203 , 1204 , 1205 , 1206 , 1207 , 1208 , 1209 , 1210 , 1211 , 1212 , 1214 , 1215 , 1216 , 1217 , 1218 , 1219 , 1220 , 1221 , 1222 , 1223 , 1224 , 1225 , 1226 , 1227 , 1228 , 1229 , 1230 , 1231 , 1232 , 1233 , 1234 , 1235 , 1236 , 1238 , 1239 , 1240 , 1241 , 1242 , 1243 , 1244 , 1245 , 1246 , 1247 , 1248 , 1250 , 1254 , 1266;   2) Symbia T2; Part Number 10275008; Serial Numbers: 1026 , 1028 , 1029 , 1035 , 1040 , 1044 , 1052 , 1054 , 1055 , 1056 , 1057 , 1058 , 1059 , 1060 , 1061 , 1062 , 1067 , 1071 , 1072 , 1073 , 1074 , 1075 , 1076 , 1077 , 1078 , 1079 , 1080 , 1081 , 1082 , 1083 , 1085 , 1086 , 1087 , 1088 , 1089 , 1090 , 1091 , 1092 , 1094 , 1095 , 1096 , 1097 , 1098 , 1099 , 1100 , 1101 , 1102 , 1103 , 1104 , 1107 , 1108 , 1109 , 1110 , 1111 , 1113 , 1114 , 1115 , 1116 , 1117 , 1118 , 1119 , 1120 , 1121 , 1122 , 1123 , 1125 , 1126 , 1127 , 1128 , 1129 , 1130 , 1131 , 1132 , 1133 , 1135 , 1136 , 1137 , 1138 , 1139 , 1140 , 1141 , 1142 , 1143 , 1144 , 1145 , 1147 , 1148 , 1149 , 1150 , 1151 , 1152 , 1153 , 1154 , 1156 , 1157 , 1158 , 1159 , 1160 , 1161 , 1162 , 1164 , 1166 , 1167 , 1168 , 1169 , 1566
510(k) numbers
["K082506"]
PMA numbers
not on file
Product code
KPS
Firm FEI
1000514697
Firm city
Hoffman Estates
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register