Skip to the content

FDA Medical Device Recalls (openFDA)

Cepheid: Brand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10

Recall number
Z-0959-2026
Recalling firm
Cepheid
Product
Brand Name: Xpert BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10
Status
Open, Classified
Initiated
2025-10-30
Posted
2025-12-18
Terminated
not on file
Reason
Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.
Root cause
Process design
Action
On November 14, 2025, Cepheid issued a "Urgent Medical Device Recall" notification to affected consignees. Cepheid asked consignees to take the following actions: 1. Cepheid recommends that laboratories and clinical facilities utilizing the Xpert BCR-ABL Ultra assay for quantification of BCR::ABL1 and ABL1 mRNA transcript levels in peripheral blood specimens from t(9;22)-positive chronic myeloid leukemia (CML) patients undergoing tyrosine kinase inhibitor (TKI) therapy, assess white blood cell count (WBCC) prior to testing in cases where elevated white blood cell count is suspected. 2. If WBCCC exceeds 30,000 cells/uL, the specimen should be processed in the accordance with the Retest Procedure for ERROR (Code 2008) or INVALID (Type 2) as detailed in the IFU for the following scenarios: - Cases where INVALID test results are observed and are associated with an elevated WBCC. - Cases of valid results in which an elevated WBCC has been observed. 3. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. 4. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them with a copy of this letter. 5. Please complete and return the enclosed Response Form within 10 days so we are assured you have received this important communication.
Quantity
6325 units
Distribution
US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe Guatemala Honduras Hong Kong India Indonesia Iraq Israel Italy Jamaica Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Luxembourg Malawi Malaysia Mauritania Mauritius Mexico Morocco Mozambique Myanmar Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pap. New Guinea Peru Philippines Poland Portugal Qatar Rep.of Congo Reunion Rwanda Saudi Arabia Senegal Seychelles Singapore South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tajikistan Tanzania Thailand Togo Tunisia Uganda United Kingdom Uruguay Utd.Arab Emir. Venezuela Zambia Zimbabwe
Lot, serial or model codes
UDI: 07332940007102/ Lot Numbers: 1001456698, 1001463946, 1001472599, 1001477457, 1001480637, 1001483326, 1001500425
510(k) numbers
["K190076"]
PMA numbers
not on file
Product code
OYX
Firm FEI
3004530258
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register