Skip to the content

FDA Medical Device Recalls (openFDA)

A L I Technologies Ltd: Change Healthcare Cardiology Hemodynamics software

Recall number
Z-0959-2025
Recalling firm
A L I Technologies Ltd
Product
Change Healthcare Cardiology Hemodynamics software
Status
Open, Classified
Initiated
2024-12-02
Posted
2025-01-16
Terminated
not on file
Reason
Software defect concerning autosave mechanism may result in the clinician administering incorrect medication(s).
Root cause
Software design
Action
On December 2, 2024 Change Healthcare Canada Company issued a "Urgent Field Safety" Recall Notice to affected consignees. Change Healthcare Canada Company ask consignees to take the following actions: 1. Please contact and work with our Customer Support team to identify and correct any potentially affected data. 2. This notice must be distributed to all personnel within your organization who need to be aware of this Urgent Field Safety Notice. Customers should alert other affiliated parties that may be affected by this Urgent Field Safety Notice. 3. To ensure effectiveness of any required corrective actions, please maintain awareness of this Urgent Field Safety Notice until the issue has been fully resolved. 4. Please notify our Customer Support department that you have read and understood this field safety notice, and if you want the product update to be installed on your system. Customer Support number by region: US / Canada: 1 877 654 4366 5. Once again, until the product update is installed at your site, Change Healthcare recommends that the immediate actions (described above) be taken.
Quantity
2 units
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, PA, SC, SD, TN, TX, WA, WI and the countries of Australia, Israel, Canada, Germany, Ireland, United Kingdom.
Lot, serial or model codes
Software version 15.01/UDI:(01)17540262100129
510(k) numbers
K230881
PMA numbers
not on file
Product code
MWI
Firm FEI
3394
Firm city
Burnaby
Firm state
not on file
Firm country
Canada

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this source