Skip to the content

FDA Medical Device Recalls (openFDA)

Getinge Usa Sales Inc: Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400

Recall number
Z-0959-2023
Recalling firm
Getinge Usa Sales Inc
Product
Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
Status
Open, Classified
Initiated
2022-12-22
Posted
2023-01-26
Terminated
not on file
Reason
Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
Root cause
Software Design Change
Action
On December 22, 2022, the firm notified affected customers through Urgent Medical Device Correction letters. Customers were instructed to examine their inventory immediately to determine if they have any of the affected serial numbers. If you have an affected system running on software version 4.8.0 through 4.8.3, you may continue use of the device until the software is updated with the following considerations: 1. If during use, the device is switched from Automatic to Manual and it is found that the manual bag is empty the user can activate the Emergency ventilation or turn off/on the device to reset the disabled valves function. 2. Users may also use the resuscitator that is available in the Operating Room. Getinge will update the software to correct the issue free of charge. You can arrange a visit by a Getinge trained or authorized service technician by contacting Getinge Customer Technical Support at (888) 9GETUSA / (888) 943 8872 (select option 1, then option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity
8,016 Total (7,766 OUS; 250 US)
Distribution
US nationwide and Worldwide Distribution: Domestic distribution to AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA, WV. Foreign distribution worldwide.
Lot, serial or model codes
Part # UDI 6677200 7325710001349 Serial Numbers 1202-8027 Part # UDI 6677300 7325710001363 Serial Numbers 1058-1201 Part # UDI 6677400 7325710001318 Serial Numbers 2034-8446 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.
510(k) numbers
["K191027"]
PMA numbers
not on file
Product code
BSZ
Firm FEI
3013876692
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register