Advanced Bionics LLC: S-Series Short Range Headpiece Model Number MMT-5300-00 (beige) and MMT-5300-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used in conjunction with speech processors manufactured by Advanced Bionics.

Recall number
Z-0959-2009
Recalling firm
Advanced Bionics LLC
Product
S-Series Short Range Headpiece Model Number MMT-5300-00 (beige) and MMT-5300-10 (brown) ADVANCED BIONICS LLC 12740 San Fernando Rd. Sylmar, CA 91342 USA. The headpiece is used in conjunction with speech processors manufactured by Advanced Bionics.
Status
Terminated
Initiated
2008-07-24
Posted
2009-02-03
Terminated
2012-03-23
Reason
Mistuned headpiece - Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Potential clinical symptoms include: -No lock -Intermittent lock -Implant locks, but cann
Root cause
Employee error
Action
Advanced Bionics is notifying clinicians of this situation so that they can provide appropriate clinical management. The notification letter sent to clinicians in the United States was sent via FedEx so that delivery information could be confirmed. The letter advises that Advanced Bionics recently became aware of a tuning issue with some Platinum Series and S-Series headpieces shipped from May 2007 through April 2008. Auria and Harmony headpieces are not affected by this issue. Tuning refers to the process through which the headpiece's communication frequency is adjusted to match that of the internal device. Since discovery of this issue firm has conducted an extensive technical analysis showing that the use of these headpieces does not pose any safety risk to the patient. While there has been no increase in the number of headpiece-related complaints, firm advises of the potential clinical symptoms: -No lock -Intermittent lock -Implant locks, but cannot measure impedances -Reduction in battery life, primarily for C1 patients with thick flaps. Per the letter, as long as the system maintains lock, there is no degradation in sound quality or stimulation. Further, the impact on battery life for CII and HiRes 90K patients is minimal. Medical professionals are advised that If they have any patients who display any of these clinical symptoms that they attribute to an affected headpiece, to advise the patient to contact Advanced Bionics Customer Service at 877-244-9541 and they will send them a replacement promptly.
Quantity
MMT-5300-10: 118 units; MMT-5300-00: 76 units
Distribution
Nationwide
Lot, serial or model codes
a) Model MMT-5300-00 (beige); Serial Numbers:14122, 14124, 14125, 14126, 14127, 14128, 14129, 14130, 14132, 14133, 14134, 14135, 14137, 14138, 14139, 14141, 14142, 14143, 14144, 14145, 14147, 14148, 14149, 14150, 14151, 14152, 14153, 14155, 14156, 14157, 14158, 14159, 14160, 14161, 14162, 14163, 14164, 14166, 14168, 14169, 14170, 14171, 14715, 14716, 14717, 14718, 14719, 14720, 14721, 14722, 14723, 14724, 14725, 14726, 14727, 14730, 14732, 14735, 14736, 14741, 14742, 14743, 14800, 14955, 14956, 14964, 14971, 15105, 15106, 15107, 15110, 15111, 15112, 15117, 15118 and 15120.   b) Model MMT-5300-10 (brown); Serial Numbers: 10250, 10251, 10252, 10254, 10255, 10256, 10289, 10290, 10291, 10292, 10293, 10294, 10295, 10296, 10297, 10298, 10299, 10300, 10301, 10302, 10303, 10304, 10305, 10306, 10307, 10308, 10309, 10310, 10311, 10312, 10313, 10314, 10315, 10316, 10317, 10318, 10319, 10320, 10321, 10322, 10323, 10324, 10325, 10326, 10327, 10328, 10329, 10330, 10331, 10332, 10333, 10334, 10335, 10336, 10337, 10338, 10339, 10340, 10341, 10342, 10344, 10345, 10346, 10348, 10352, 10354, 10362, 10367, 10374, 10375, 10378, 10559, 10561, 10566, 10572, 10573, 10577, 10578, 10582, 10583, 10586, 10587, 10588, 10649, 10650, 10651, 10652, 10653, 10654, 10655, 10656, 10657, 10658, 10659, 10660, 10661, 10663, 10664, 10665, 10666, 10668, 10669, 10670, 10673, 10674, 10675, 10676, 10677, 10678, 10682, 10684, 10702, 10705, 10769, 10770, 10771, 10773 and 10774.
510(k) numbers
not on file
PMA numbers
["P960058"]
Product code
MCM
Firm FEI
3006556115
Firm city
Sylmar
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register