Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc: Cytomics FC 500 MPL Flow Cytometry System with CXP Software, Software Version 2.2, Automated differential cell counter, Part Number 626553, Beckman Coulter, Fullerton, CA 92834

Recall number
Z-0959-2008
Recalling firm
Beckman Coulter Inc
Product
Cytomics FC 500 MPL Flow Cytometry System with CXP Software, Software Version 2.2, Automated differential cell counter, Part Number 626553, Beckman Coulter, Fullerton, CA 92834
Status
Terminated
Initiated
2007-06-29
Posted
2008-04-30
Terminated
2012-12-13
Reason
Data generation error: If the cytosettings is not refreshed or restarted during data importation, the output will include old and new data.
Root cause
Software design
Action
Consignees were notified by an Urgent Product Corrective Action letter sent on 6/29/07. The letter provided details on the data error and instruction on how to avoid the error. Users were asked to retain notification as part of the Quality Systems documentation and return a response form acknowledging receipt. For additional information, contact 1-800-526-7694.
Quantity
516 units in the US; 73 units in Canada
Distribution
Worldwide Distribution- USA and country of Canada.
Lot, serial or model codes
Serial Numbers: AL11058, AK20149, AJ37243, AJ08035, AK16130, AK01004, AJ33215, AK16129, AK25182, AK51355, AL13064, AJ38253, AK51353, AK07059, AL12062, AJ40271, AJ02008, AJ50371, AK07058, AL21134, AJ01002, AK12093, AL15085, AL14077, AJ30191, AJ48342, AK20152, AK07060, AJ37246, AJ06026, AK32539, AJ39259, AL04022, AJ37240, AK48336, AJ46318, AJ14074, AJ17095, AK34227, AJ47331, AJ48341, AJ33210, AK48347, AK20158, AK12095, AK16131, AK19146, AK17138, AJ25164, AJ24147, AJ51381, AJ48339, AK51359, AJ49351, AL14082, AL22140, AJ24155, AJ47332, AK51358, AL23144, AJ24151, AJ49348, AK51380, AJ38250, AK48343, AK17135, AK18141, AK09075, AK13103, AJ29184, AJ30189, AJ02006, AJ14075, AJ14073, AJ15079, AJ16087, AJ16085, AJ15078, AJ16083, AJ19103, AJ19111, AJ21116, AJ19109, AJ21114, AJ21115, AJ21117, AJ22121, AJ22122, AJ23135, AJ25165, AJ24156, AJ25159, AJ24154, AJ25167, AJ25161, AJ47327, AK07062, AK33215, AL18108, AL20125, AJ17091, AK34232, AJ25160, AJ25163, AK05025, AJ29181, AJ50366, AK48328, AJ44313, AK51370, AK34230, AJ38248, AL19111, AJ12063, AJ44309, AJ22124, AK08071, AK48334, AK25177, AJ48344, AJ42294, AK51372, AL12063, AJ24158, AJ46319, AJ43298, AK30199, AK20153, AK51354, AJ31195, AJ49349, AJ12064, AK33225, AJ35229, AK17140, AJ29183, AK16127, AJ36234, AK17134, AL09044, AJ48340, AK19147, AK39264, AL21133, AL06033, AJ19110, AJ10048, AK51361, AJ24148, AJ50368, AJ11056, AJ34221, AJ03012, AK19148, AK06051, AK34233, AK39246, AJ14072, AJ44316, AJ42292, AK48338, AK34229, AJ29182, AK18139, AK08066, AK44284, AJ05020, AK30204, AK08063, AJ34217, AJ23129, AJ23134, AJ34226, AJ37238, AK25181, AK34231, AK34228, AK34234, AK48342, AJ24142, AK39266, AJ47326, AL21135, AJ07032, AL16093, AJ38251, AJ38252, AJ43300, OAK48341, AJ23138, AJ25166, AL23145, AK39273, AJ30194, AK02009, AJ24144, AJ32209, AL13073.
510(k) numbers
["K030828"]
PMA numbers
not on file
Product code
GKZ
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register