ICU Medical, Inc.: 154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).

Recall number
Z-0958-2021
Recalling firm
ICU Medical, Inc.
Product
154 cm (61 IN) APPX 1.5ml, PUR YELLOW EXT, 1.2 MICRON FILTER - Product Usage: The intravascular administration set is a single use, sterile, non pyrogenic device, which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle or catheter (which is inserted into a vein).
Status
Terminated
Initiated
2020-12-02
Posted
not on file
Terminated
2023-06-30
Reason
Due to the incorrect filter assembly.
Root cause
Process control
Action
On 12/02/2020, the firm's third party contractor send an "URGENT: MEDIAL DEVICE RECALL" Notification letter via traceable mailing to direct and indirect consignees informing them that specific lots of IV extension sets that may contain an incorrect filter assembly of a 0.2 micron filter instead of a 1.2 micron filter and the potential hazards/risks potentially cause delay in therapy or under delivery of medication. Customers are instructed to: 1) Discontinue the use and distribution of the affected product immediately. Check inventory and quarantine all affected product at their facility. 2) Inform potential users of the product in their organization of the notification and to complete the attached response form. Return the completed response form to the fax number or e-mail address on the form, even if customer do not have the affected product. 3) Return affected product using the return label provided with the letter. Contact Stericycle at 1-888-243-5547 (MF, 8am-5pm ET) if they have not received a return label or require additional labels for returning the affected product. The return labels are for single use only. Please do not reproduce. Visit http://expertezlabel.com to request additional labels for returning affected product. To ensure proper and timely credit, follow the instructions on the return label for returning product. Upon receipt of the completed response form and return of the affected product, the Recalling Firm will credit you for any product returned. Customers will only receive credit for product that they return. NOTE: Credits for product purchased through distributor will be credited by the distributor. 4) If customers have distributed the product further, immediately notify the accounts that received the product identified in the Affected Product section of this notification and ask them to contact Stericycle at 1-888-243-5547 (M-F, 8am-5pm ET) to obtain a response form. For further inquiries, please contact the Recalling Firm usi
Quantity
Total Kits = 7,500 for all four lots
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AZ, CA, CO, CT, FL, ID, IL, LA, MA, MN, NC, NJ, NY, TN, TX, VA, WA and WV. The country of Belgium.
Lot, serial or model codes
Distributed O.U.S. only Model List Number: 011-H3182 UDI: (01)1084061905261(17)250401(30)50(10)4806428 Lot Number: 4806428
510(k) numbers
["K964435"]
PMA numbers
not on file
Product code
FMG
Firm FEI
2025816
Firm city
San Clemente
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register