Medtronic Inc.: Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution

Recall number
Z-0958-2020
Recalling firm
Medtronic Inc.
Product
Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution
Status
Open, Classified
Initiated
2019-11-21
Posted
2020-02-07
Terminated
not on file
Reason
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
Root cause
Device Design
Action
On November 21, 2019, Medtronic initiated a recall of MiniMed 600 series insulin pumps with a clear retainer ring design. On October 5, 2021, the firm expanded the recall and issued an "Urgent Field Safety Notification" to all patients and health care providers. The updated notification informs consignees that Medtronic will proactively replace all MiniMed 600 series insulin pumps that have a clear retainer ring design with an equivalent pump that has the updated black retainer ring design. Pumps will be replaced at no charge over the coming months even if the retainer ring is not damaged and regardless of the warranty status of the pump. In addition to providing consignees with this updated recall strategy, the firm asked consignees to take the following actions: ACTIONS REQUIRED BY YOU: 1. Complete and return the attached Confirmation Form to acknowledge that you have reviewed and understood this notification. 2. Medtronic is requesting patients visit [www.medtronicdiabetes.com/PumpRing] and complete the form online, or call us at 1-877-585-0166 to indicate their decision to receive a replacement pump at no charge. Replacement pumps will become available in the coming months and patients will be notified when their pump is ready to ship. Please assist patients in making replacement decision considering the safety risk associated with the retainer ring issue. For additional recall details, please view the following link: https://www.medtronicdiabetes.com/customer-support/product-and-service-updates/notice15-letter
Quantity
not on file
Distribution
US:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Republic, Finland, Switzerland, Slovakia, Ireland, Poland, S. Africa, Greece, Hungary, Saudi Arabia, United Arab Emira, New Zealand, Australia, Japan, Mexico, Colombia, Brazil, S. East Asia, Korea, Canada,
Lot, serial or model codes
All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US***
510(k) numbers
not on file
PMA numbers
not on file
Product code
OZP
Firm FEI
3003166194
Firm city
Northridge
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register