Skip to the content

FDA Medical Device Recalls (openFDA)

Invivo Corporation: Philips Expression MR200 MRI Patient Monitoring System.

Recall number
Z-0958-2015
Recalling firm
Invivo Corporation
Product
Philips Expression MR200 MRI Patient Monitoring System.
Status
Terminated
Initiated
2014-10-22
Posted
2015-01-13
Terminated
2017-01-09
Reason
The MR 200 devices failed the initial power up test due to capacitor short and hi-pot testing; possible loss of Patient Monitoring or User/Service Electrical Shock.
Root cause
Component design/selection
Action
Philips Medical Systems sent an Urgent - Medical Device Correction Letter dated October 22, 2014, to all affected customers. The letter identified the product, the problem and the action needed to be taken by the customer. Our records indicate that you have an affected device. The following page provides additional instructions and actions that will be taken to address this problem. If you need any further information or support concerning this issue, please contact your local Philips representative: 877-468-4861 option 1, then option 2.
Quantity
20
Distribution
CA, FL, IN, KY, MN, NC, NJ, NY, OK, SD and TX Netherlands, Poland and United Kingdom.
Lot, serial or model codes
Model No. 866120; Lot/ Serial Numbers: US33600065, US33600066, US33600075, US33600077, US33600078, US33600079; US33600080, US33600081, US33600082, US33600083, US33600084, US33600086, US33600087, US33600088, US33600089, US33600090, US33600091, US33600092, US33600093, US33600094.
510(k) numbers
["K131382"]
PMA numbers
not on file
Product code
MWI
Firm FEI
1051786
Firm city
Orlando
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register