Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott: Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile.

Recall number
Z-0957-2022
Recalling firm
Abbott
Product
Abbott SE TactiCath Sensor Enabled Contact Force Ablation Catheter, Bi-D, Curve DD, Model A-TCSE-DD, Sterile.
Status
Open, Classified
Initiated
2022-03-24
Posted
not on file
Terminated
not on file
Reason
Devices were manufactured with a "First Use Date" preprogrammed in the EEPROM which disables the use of the single-use device.
Root cause
Process control
Action
The recalling firm issued letters dated 3/24/2022 on 3/24/22 via hand-delivery. The letters explained the issue, the impact and associated risks, and the steps to reduce risk. The steps included not using any of the remaining inventory of the affected lots listed in Appendix A, complete and return the accompanying Acknowledgment Form, and return all remaining unused affected devices to the recalling firm. The consignee is informed their Abbott representative can assist in returning these devices and obtaining replacements. The notice is to be forwarded to anyone within their organization who may need to be notified and a copy of the notice should be kept along with the completed Acknowledgment Form for their records.
Quantity
588 devices
Distribution
Worldwide distribution - US Nationwide - there was government distribution but no military. The countries of Austria, Belgium, Canada, Denmark, France, Germany, Greece, Israel, Italy, Japan, Netherlands, Norway, Palestine, Poland, Portugal, Russia, Slovakia, Spain, Sweden, Taiwan, and United Kingdom.
Lot, serial or model codes
Lot numbers: 8181068, 8182131, 8198290, 8198350, 8200249, 8200300, 8201447, 8202313, 8203518, 8203835, 8203857, 8205760, and 8207547. UDI (01)05415067027634(17)230331(10)xxxxxxx (lot number).
510(k) numbers
not on file
PMA numbers
["P130026"]
Product code
OAE
Firm FEI
3005334138
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register