FDA Medical Device Recalls (openFDA)
Myofunctional Research Company USA: Myosa for Kids KS1 REF MyOSAKS1_Sb/Sp/Mb/Mp/Lb/Lc KS2 REF MyOSAKS2_Sb/Sc/Mb/Mc/Mp/Lb/Lc Myosa for Kids is a mouthguard in 3 different sizes (sm, med, lg) that can be used in children ages 3-15 to address breathing and myofunctional disorders. Catalog Numbers:
- Recall number
- Z-0956-2026
- Recalling firm
- Myofunctional Research Company USA
- Product
- Myosa for Kids KS1 REF MyOSAKS1_Sb/Sp/Mb/Mp/Lb/Lc KS2 REF MyOSAKS2_Sb/Sc/Mb/Mc/Mp/Lb/Lc Myosa for Kids is a mouthguard in 3 different sizes (sm, med, lg) that can be used in children ages 3-15 to address breathing and myofunctional disorders. Catalog Numbers:
- Status
- Open, Classified
- Initiated
- 2025-10-09
- Posted
- 2025-12-18
- Terminated
- not on file
- Reason
- Due to firm mistakenly distributing product within the U.S. without FDA premarket clearance/approval.
- Root cause
- No Marketing Application
- Action
- On 10/09/2025, the firm emailed an "URGENT: Medical Device Recall" letter to customers informing them that products (Myosa for Kids KS1 and Myosa for Kids KS2) were distributed in the United States without the required FDA premarket clearance. Customers are instructed to: 1. Immediately examine their inventory for the affected products. 2. Cease distribution/use of any remaining Myosa KS1 and KS2 devices. 3. Quarantine affected stock to prevent further use. 4. Complete and return the enclosed Response Form within 10 business days, even if customers no longer have any affected product. 5. Return quarantine devices using the instructions provided after receipt of customer response forms For any patient who has initiated treatment with the affected devices or to whom you have not explained the risks associated with these devices, it is requested that you notify them and provide warning s and clinical alternatives as appropriate. For questions and/or assistance - Contact Recall Coordinator at 909-587-4940 or email usa.hq@myoresearch.com
- Quantity
- 436 devices
- Distribution
- U.S. (nationwide) distribution to states of: Nationwide: AK, AR, CA, CO, FL, GA, HI, IL, KS, KY, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OR, PA, TX, UT, VA, and WA; O.U.S. (foreign) to country of: Australia
- Lot, serial or model codes
- Myosa for Kids KS1 REF MyOSAKS1_Sb/Sp/Mb/Mp/Lb/Lc -S1 Large Blue -S1 Large Clear -S1 Medium Blue -S1 Medium Pink -S1 Small Blue -S1 Small Pink KS2 REF MyOSAKS2_Sb/Sc/Mb/Mc/Mp/Lb/Lc -Stage 2 Large Blue -Stage 2 Large Clear -Stage 2 Medium Blue -Stage 2 Medium Clear -Stage 2 Medium Pink -Stage 2 Small Blue -Stage 2 Small Clear
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LRK
- Firm FEI
- 3005505603
- Firm city
- Rancho Cucamonga
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28