Siemens Medical Solutions USA, Inc: ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor
- Recall number
- Z-0956-2023
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor
- Status
- Open, Classified
- Initiated
- 2022-12-16
- Posted
- 2023-01-13
- Terminated
- not on file
- Reason
- Mechanical connection between the tabletop and table base may be lost.
- Root cause
- Device Design
- Action
- A customer letter dated 12/16/2022 was issued to affected consignees. Consignees are advised the following: -Before treatment is started, check if both transversal guide rails and the roller bearing at the end of each rail are in the correct position. -For the affected OR tables, an inspection including installation of a safety measure will be carried out. Siemens is currently developing this mitigation. Once correction is available, the service department will contact the consignees to arrange a date to perform the corrective action. Consignees are requested to acknowledge the recall letter and return the acknowledgement form to Siemens promptly.
- Quantity
- 550 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Italy, Japan, Kazakhstan, Kuwait, Lebanon, Monaco, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, and Vietnam.
- Lot, serial or model codes
- 1) ARTIS pheno; Material #10849000; UDI/DI: 4056869046877 2) ARTIS icono biplane; Material #11327600; UDI/DI: 4056869063317 3) ARTIS icono floor; Material #11327700; UDI/DI: 4056869149325
- 510(k) numbers
- ["K163286","K190768"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 2240869
- Firm city
- Malvern
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28