Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc: ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor

Recall number
Z-0956-2023
Recalling firm
Siemens Medical Solutions USA, Inc
Product
ARTIS pheno, ARTIS icono biplane, and ARTIS icono floor
Status
Open, Classified
Initiated
2022-12-16
Posted
2023-01-13
Terminated
not on file
Reason
Mechanical connection between the tabletop and table base may be lost.
Root cause
Device Design
Action
A customer letter dated 12/16/2022 was issued to affected consignees. Consignees are advised the following: -Before treatment is started, check if both transversal guide rails and the roller bearing at the end of each rail are in the correct position. -For the affected OR tables, an inspection including installation of a safety measure will be carried out. Siemens is currently developing this mitigation. Once correction is available, the service department will contact the consignees to arrange a date to perform the corrective action. Consignees are requested to acknowledge the recall letter and return the acknowledgement form to Siemens promptly.
Quantity
550 units
Distribution
Worldwide distribution - US Nationwide and the countries of Armenia, Australia, Austria, Belgium, Bulgaria, Canada, Chile, China, Colombia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Indonesia, Iran, Ireland, Italy, Japan, Kazakhstan, Kuwait, Lebanon, Monaco, Netherlands, New Zealand, Norway, Oman, Poland, Qatar, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, and Vietnam.
Lot, serial or model codes
1) ARTIS pheno; Material #10849000; UDI/DI: 4056869046877 2) ARTIS icono biplane; Material #11327600; UDI/DI: 4056869063317 3) ARTIS icono floor; Material #11327700; UDI/DI: 4056869149325
510(k) numbers
["K163286","K190768"]
PMA numbers
not on file
Product code
OWB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register