Skip to the content

FDA Medical Device Recalls (openFDA)

Synthes USA HQ, Inc.: Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.

Recall number
Z-0956-2013
Recalling firm
Synthes USA HQ, Inc.
Product
Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.
Status
Terminated
Initiated
2013-01-11
Posted
2013-03-15
Terminated
2015-08-31
Reason
It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
Root cause
Labeling mix-ups
Action
The firm initiated their recall of this product on January 11, 2013 by sending an Urgent Notice: Medical Device Recall letter to consignees. The letter identified the affected product and the issue. Synthes requested customers to examine their inventory for the affected product, remove them from use, and return them to Synthes. Additionally, customers were to follow the steps provided in the event that they did or did not have any affected product on hand. Questions should be directed to 610-719-5450 or FieldAction@synthes.com.
Quantity
4
Distribution
Distributed in the states of IN, OH, PA, and TX.
Lot, serial or model codes
Lot number 7965181
510(k) numbers
["K102694"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3005180112
Firm city
West Chester
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register