Skip to the content

FDA Medical Device Recalls (openFDA)

Roche Diagnostics Corp.: MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.

Recall number
Z-0956-2007
Recalling firm
Roche Diagnostics Corp.
Product
MagNA Pure LC Total Nucleic Acid Isolation Kit - Large Volume; Catalog # 03264793001.
Status
Terminated
Initiated
2007-05-14
Posted
2007-06-22
Terminated
2009-12-17
Reason
False Positive Results: Product labeling states cross contamination is not an issue for high viral load titer samples, but the firm has determined that a potential sample contamination may not be detected at these levels and could result in false positive results being reported.
Root cause
Other
Action
Consignees were sent an Urgen Medical Device Correction letter dated 5/14/07 which instructs them to define and establish a lower detection limit, to use UNG to prevent carryover whenever possible, and to confirm low positive results through an independent experiment. A revised device correction letter dated 6/15/07 was subsequently sent to each consignee updating the reason for recall and associated risks.
Quantity
4,201
Distribution
Nationwide.
Lot, serial or model codes
Lots 11962700, exp. 8/31/07; 12030400, exp. 6/30/07; 12071100, exp. 9/30/07; 12125800, exp. 10/31/07; 12214300, exp. 12/31/07; 12360200, exp. 3/31/08; 12426200, exp. 4/3008; 12543600, exp. 5/31/08; 12672500, exp. 5/31/08; 12725300, exp. 6/30/08; 12799800, exp. 7/31/08; 12917900, exp. 8/31/08; 13024000, exp. 9/30/08; 13081700, exp. 9/30/08; 13197600, exp. 12/31/08 and 13266600, exp. 2/28/09.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NPQ
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register