FDA Medical Device Recalls (openFDA)
Boston Scientific CRM Corp: VITALITY AVT, model A135 and model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Recall number
- Z-0955-05
- Recalling firm
- Boston Scientific CRM Corp
- Product
- VITALITY AVT, model A135 and model A155. Implantable Cardioverter Defibrillator. Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
- Status
- Terminated
- Initiated
- 2005-06-17
- Posted
- 2005-06-30
- Terminated
- 2008-03-02
- Reason
- The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.
- Root cause
- Other
- Action
- Firm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.
- Quantity
- 22,575 devices (19,446 implanted world wide)
- Distribution
- worldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Syrian Arab Republic, Turkey, US - Virgin Islands and United Kingdom.
- Lot, serial or model codes
- all serial numbers
- 510(k) numbers
- not on file
- PMA numbers
- ["P960040/S29","P960040S035"]
- Product code
- LWP
- Firm FEI
- 2124215
- Firm city
- Saint Paul
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as