Skip to the content

FDA Medical Device Recalls (openFDA)

CoreLink LLC: CoreLink LLC, FLXFIT 15, 32MM X 8MM, CAGE, REF: AT3208

Recall number
Z-0954-2023
Recalling firm
CoreLink LLC
Product
CoreLink LLC, FLXFIT 15, 32MM X 8MM, CAGE, REF: AT3208
Status
Terminated
Initiated
2022-10-20
Posted
2023-01-12
Terminated
2024-08-20
Reason
Specific lots of FLXfit implants have internal components that may have been manufactured using stainless steel rather than the specified titanium alloy.
Root cause
Process control
Action
CoreLink initiated their recall by telephone to the physicians on 10/20/2022. The firm followed with a letter sent by FedEx. The letter explained the problem and the potential risk for health: "Potential Risks to Health: Please be aware that our records indicate your practice may have used one or more of the affected AT3208 devices (as shown in the table above), and that patients implanted with a FLXfit Expandable Interbody Spacer from affected lots should be monitored for: 1. MRI Safety: FLXfit devices, per current labeling, are not approved in the MR environment, and additionally, this labeling does not specify the presence of 17-4 Surgical Grade Stainless Steel. There is a possible risk of patient injury while scanning a patient implanted with FLXfit devices. The additional risk from the presence of steel is believed to be very low as published studies1,2,3, have shown negligible implant heating and no measurable device migration in an MR environment at 3 Tesla with stainless steel implants containing as much as 65% Iron content by weight and 31 times great mass than the affected devices1. 2. Implant Corrosion: There is a remote possibility for implant corrosion over time, which may result in localized pain, swelling, and tenderness, and may weaken the implant. If fusion does not occur, there is an increased potential for implant failure. CoreLink believes this is a remote possibility due in part to the passivation process performed in accordance with ASTM A967 that removes free iron particles on the surface of the affected component and creates a corrosion resistant passive surface layer. 3. Nickel Sensitivity: Patients with nickel sensitivities may experience non-life threatening4,5 allergic reactions to the implant material and implant removal may be necessary to alleviate these effects. We do not anticipate this issue to affect the structural integrity of the implant. The level of Nickel in the component by weight is .9% of the total mass, this is compara
Quantity
18 units
Distribution
US Nationwide distribution.
Lot, serial or model codes
UDI/DI M725AT32080, Lot Codes: SM135217, SM129260
510(k) numbers
["K171519"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3006494201
Firm city
Fenton
Firm state
MO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register