Covidien LP: Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien, Valleylab, a division of Tyco Healthcare Group LP, 5920 Longbow Drive, Boulder, CO 80301 Intended use: LigaSure 5 mm devices are electrosurgical instruments designed to permanently seal and divide vessels, tissue bundles, and lymphatics.

Recall number
Z-0954-2011
Recalling firm
Covidien LP
Product
Covidien, LigaSure, Blunt Tip Laparoscopic Sealer/Divider, 5 mm-37 cm, REF LF1537, Sterile EO, Covidien, Valleylab, a division of Tyco Healthcare Group LP, 5920 Longbow Drive, Boulder, CO 80301 Intended use: LigaSure 5 mm devices are electrosurgical instruments designed to permanently seal and divide vessels, tissue bundles, and lymphatics.
Status
Terminated
Initiated
2010-12-08
Posted
2011-01-25
Terminated
2012-03-27
Reason
During surgical procedures, the device may energize when clamping without pushing the activation button, which could result in unintended energy delivery to tissue.
Root cause
Device Design
Action
The firm, Covidien, sent a "URGENT: PRODUCT RECALL" letter dated December 6, 2010, to the customers. The letter described the product, problem and the action to be taken by the customers. The customers were instructed to remove product from inventory, immediately quarantine the product and return any remaining stock of the lots listed to: Covidien Energy-based devices, Attention: R82, Building 4/Receiving, 5920 Longbow Drive, Boulder, CO 80301. The customers were also instructed to complete and return the Customer Response Form immediately via mail (in the envelope provided) or fax to (303) 516-8307. To return any identified product for credit or replacement, contact Customer Service at 1-800-722-8772 Option 1. Should you have any medical questions concerning this recall, please contact your local sales representative or the Covidien Clinical information Hotline at (800) 255-8522 (option 1).
Quantity
25,009 units
Distribution
Worldwide distribution: USA and countries including: Australia, Canada, France, Germany, Japan, Puerto Rico, Singapore, and Uruguay.
Lot, serial or model codes
Lot No.: 188856, 190048, 190107, and 190307 through 193333 (88 lots).
510(k) numbers
["K092879"]
PMA numbers
not on file
Product code
GEI
Firm FEI
1717344
Firm city
Boulder
Firm state
CO
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register