Millipore Corporation: LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit.

Recall number
Z-0954-2009
Recalling firm
Millipore Corporation
Product
LIGHT DIAGNOSTICS Simulfluor Flu A / Flu B 2 mL Reagent Catalog # 5250 which is a component in the Kit Catalog #3121 Simulfluor Flu A / Flu B DFA kit.
Status
Terminated
Initiated
2008-11-24
Posted
2009-03-06
Terminated
2012-08-10
Reason
Certain lots of Simulfluor Flu A / Flu B Reagent may cross-react with Respiratory Synctial Virus. The continued use of Simulfluor Flu A / Flu B Reagent Catalog #5250 may lead to a potential misdiagnosis of influenza B.
Root cause
Pending
Action
Millipore is performing a voluntary recall by notifying customers and distributors by telephone on November 24, 2008. In addition, an Urgent: Medical Device Recall notification letter with the same date was sent via certified mail. Customers were advised to examine their stock immediately and discontinue the use of the specified lots. Customers were instructed to complete the enclosed Effective Check Survey confirming disposal of defective lots and return this form to Millipore's Regulatory Department. Customer service is to be contacted at 1-800-437-7500 for replacement product. Millipore Corporation will issue credit or replace the recalled lots that customers currently have in inventory.
Quantity
258 units
Distribution
Worldwide Distribution -- USA including states of AL, CA, CO, CT, FL, ID, IL, KS, MA, MI, MS, NM, NY, OH, OK, OR, PA, SD, TX, UT, VA, VT, WA, WV, and WI and countries of Canada, Mexico, Italy, Belgium,Germany, France, Spain, and United Kingdom.
Lot, serial or model codes
Lot Numbers JBC1531414, TR1458084, TR1480746, JBC1531421, JH1534500, TR1501923, TR1417106, and TR1422506.
510(k) numbers
["K974302"]
PMA numbers
not on file
Product code
GNW
Firm FEI
2024094
Firm city
Temecula
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register