FDA Medical Device Recalls (openFDA)

Boston Scientific CRM Corp: VENTAK PRIZM AVT model no. 1900, Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.

Recall number
Z-0954-05
Recalling firm
Boston Scientific CRM Corp
Product
VENTAK PRIZM AVT model no. 1900, Atrial and Ventricular Therapies. Sterilized using gaseous ethylene oxide. Guidant Corporation Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798, USA.
Status
Terminated
Initiated
2005-06-17
Posted
2005-06-30
Terminated
2008-03-02
Reason
The atrial therapy (AVT) subgroups of certain Guidant ICD and CRT-D product families are subject to a condition in which a random memory error causes functional 'latching' that limits available therapy.
Root cause
Other
Action
Firm issued a Dr. Doctor letter dated 06/17/05. Letter informs physicians of the issue, clinical implications, recommendations and action taken by the firm.
Quantity
937 devices (714 currently implanted worldwide)
Distribution
worldwide. To the US and OUS countries to include: Argentina, Australia, Austria, Belgium, Canada, Chile, Cyprus, Denmark, France, Germany, Guadeloupe, Hong Kong, Ireland, Israel, Italy, Japan, Jordan, Lebanon, Luxembourg, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, Syrian Arab Republic, Turkey, US - Virgin Islands and United Kingdom.
Lot, serial or model codes
all serial numbers
510(k) numbers
not on file
PMA numbers
["P960040S028"]
Product code
LWP
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register