In2Bones, SAS: SCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW Plating System is intended for fixation of intra-articular and extraarticular fractures of the distal radius and reconstruction of the distal radius

Recall number
Z-0952-2017
Recalling firm
In2Bones, SAS
Product
SCREW 02.7-12MM LOCKING, REF W27 ST112: a) DRP, b) NEO, In2Bones, 28 chemin du petit bois, 69130 Ecully - France The NEOVIEW Plating System is intended for fixation of intra-articular and extraarticular fractures of the distal radius and reconstruction of the distal radius
Status
Terminated
Initiated
2016-12-19
Posted
2017-01-06
Terminated
2017-02-01
Reason
Dimensional non-conformity of the thread on the screw head.
Root cause
Component design/selection
Action
IN2BONES en a Field Safety Corrective Action letter dated December 15, 2016, to the sole US distributor. The letter identified the product, the problem, and the action to be taken by the customer. We therefore recommend you to follow the instructions here below: 1. Identify all the devices with the batches subject of this recall that might still be in your inventory and quarantine them 2. Inform and distribute this Recall Notification to all relevant persons within your organization 3. To distributors only: Identify all the devices with the batches subject of this recall that were delivered to your customers, and if relevant, instruct them to also follow these instructions (identification and quarantine) 4. Fill in and return the fax back form enclosed. With this form, you will certify that you have received this Recall Notification and intend to comply with the recommendations listed. This fax back form will enable In2Bones to conduct effectiveness checks. Customers with questions were instructed to contact the Quality and Regulatory Affairs team at +33 4 72 29 26 26 / +33 7 61 88 44 30 or by email: qualte@in2bones.com. For questions regarding this recall call 844-602-6637.
Quantity
19 units
Distribution
Worldwide Distribution - US TN only and Internationally to France and Switzerland
Lot, serial or model codes
Lot Numbers: a) 1602064, b) 1606067
510(k) numbers
["K160995"]
PMA numbers
not on file
Product code
HRS
Firm FEI
3010470577
Firm city
Ecully
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register