Skip to the content

FDA Medical Device Recalls (openFDA)

Depuy Ireland UC: BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR Catalog Numbers: 546308000 (Size 8); 546310000 (Size 10); 546312000 (Size 12); 546314000 (Size 14); 546316000 (Size 16); 546318000 (Size 18); 546320000 (Size 20)

Recall number
Z-0951-2023
Recalling firm
Depuy Ireland UC
Product
BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR Catalog Numbers: 546308000 (Size 8); 546310000 (Size 10); 546312000 (Size 12); 546314000 (Size 14); 546316000 (Size 16); 546318000 (Size 18); 546320000 (Size 20)
Status
Open, Classified
Initiated
2022-12-19
Posted
2023-01-12
Terminated
not on file
Reason
All lots of BIOSTOP G Bioresorbable Cement Restrictor are being removed as a precautionary measure because tested endotoxin levels were higher than recommended by the current FDA regulatory guidance. In general, endotoxins have a potential to initiate inflammatory responses, ranging from a mild fever to potentially impact or damage to vital organs.
Root cause
Process control
Action
On December 19, 2022, the firm notified affected consignees through "Urgent Medical Device Recall" letters. Customers were instructed to immediately identify and quarantine affected product in inventory. Recalled products should not be used. Customers should then contact Sedgwick at synthes5716@sedgwick.com or call 888-597-5387 to coordinate the return/credits of the subject products.
Quantity
2199 US; 43515 OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Croatia, Czech Republic, Denmark, Finland, France, Germany, Great Britain, Hungary, Ireland, Israel, Italy, Latvia, Lebanon, Luxemburg, Mexico, Netherlands, New Zealand, Norway, Portugal, Reunion, Romania, Russia, Slovakia, South Africa, Spain, Sweden, Switzerland, and Uruguay.
Lot, serial or model codes
Catalog Number Part Description UDI/DI * 546308000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 8 10603295174547 546310000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 10 10603295174554 546312000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 12 10603295174561 546314000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 14 10603295174578 546316000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 16 10603295174585 546318000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 18 10603295174592 546320000 BIOSTOP G BIORESORBABLE CEMENT RESTRICTOR SIZE 20 10603295174608 * UDI-DI on GS1 0603295a0035792 Expiry dates December 2022 thru May 2025 ALL LOTS
510(k) numbers
["K943727"]
PMA numbers
not on file
Product code
LZN
Firm FEI
3015516266
Firm city
Ringaskiddy Co. Cork
Firm state
Munster
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register