FDA Medical Device Recalls (openFDA)
Arrow International Inc: AF 550x8ML/H 2ML 15LO 8.7" DUAL CATHS, 4, Product code MT8015XLYK10 CP, Arrow AutoFuser Disposable Pain Control Pump
- Recall number
- Z-0951-2019
- Recalling firm
- Arrow International Inc
- Product
- AF 550x8ML/H 2ML 15LO 8.7" DUAL CATHS, 4, Product code MT8015XLYK10 CP, Arrow AutoFuser Disposable Pain Control Pump
- Status
- Terminated
- Initiated
- 2018-10-02
- Posted
- not on file
- Terminated
- 2021-11-12
- Reason
- Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
- Root cause
- Process control
- Action
- On October 2, 2018, the firm notified consignees via an Urgent Medical Device Recall letter. The letter informed customers of the product issue. Customers were instructed to take the following actions: 1. If you have affected stock in inventory, immediately discontinue use and quarantine the products. 2. Please complete the enclosed Recall Acknowledgement Form as soon as possible indicating whether you do or do not have stock and fax it to 1-855-419-8507 or email to recalls@teleflex.com. A customer service representative will contact you with a Return Goods Authorization (RGA) number and provide instructions for the return of products. If you have any other questions, feel free to contact your local sales representative or Teleflex Customer Service at 1-866-396-2111.
- Quantity
- 8
- Distribution
- Nationwide distribution.
- Lot, serial or model codes
- 74A1701382 74H1701147 74B1602450 74J1600057 74B1701726 74L1601123 74F1701975 74L1702110
- 510(k) numbers
- ["K090300"]
- PMA numbers
- not on file
- Product code
- MEB
- Firm FEI
- 3015859709
- Firm city
- Reading
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28