Diagnostica Stago, Inc.: Vial Label: 1) LIATEST VWF:Ag Buffer - 5 ml Diagnostica Stago 2) LIATEST VWF:Ag Latex + 4 ml R3 Diagnostica Stago 3) LIATEST VWF:Ag Latex Diluent 4 mL Diagnostica Stago Model, catalogue, number 00518 - 58978 - 58979 Kit Label: LIATEST VWF:Ag DIAGNOSTICA STAGO S.A.S. 92600 Asnires sur Seine (France) Coagulation Assay

Recall number
Z-0951-2013
Recalling firm
Diagnostica Stago, Inc.
Product
Vial Label: 1) LIATEST VWF:Ag Buffer - 5 ml Diagnostica Stago 2) LIATEST VWF:Ag Latex + 4 ml R3 Diagnostica Stago 3) LIATEST VWF:Ag Latex Diluent 4 mL Diagnostica Stago Model, catalogue, number 00518 - 58978 - 58979 Kit Label: LIATEST VWF:Ag DIAGNOSTICA STAGO S.A.S. 92600 Asnires sur Seine (France) Coagulation Assay
Status
Terminated
Initiated
2013-01-28
Posted
2013-03-14
Terminated
2013-07-19
Reason
Diagnostica Stago has confirmed that sometimes the delta OD measured can be changed leading to a lowered percentage result.
Root cause
Software design (manufacturing process)
Action
Diagnostica Stago Inc. sent e-mail notifications on 1/28/2013 to all consignees (laboratory supervisor). The written field safety notice, entitled Important Field Safety Notice, with its response form was attached. The letter identified the affected product and gave a description of the defect. Customers were asked to confirm all the results obtained in the abnormal range by either of the two methods provided. Users are to complete and return the enclosed response form as soon as possible to the Stago Hotline via fax, 973-644-9348. Questions should be directed to the Stago Hotline at 800-725-0607.
Quantity
2901 kits
Distribution
Worldwide Distribution - USA and Canada.
Lot, serial or model codes
Lot/Unit Numbers : Combination of All STA-R¿/ STA-R EVOLUTION¿ and lots of STA¿ - LIATEST¿ VWF: AG are involved.    Lot Number Expiry date 108128 2013-03 108512 2013-04 108778 2013-07 109415 2013-10 109591 2014-01
510(k) numbers
["K962675"]
PMA numbers
not on file
Product code
GJT
Firm FEI
2245451
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register