Skip to the content

FDA Medical Device Recalls (openFDA)

Guidant Corporation: CONTAK RENEWAL 4 CRT-D (model no. H190 and H195) and CONTAK RENEWAL 4 HE CRT-D (model no. H197 and H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 5511-5798 USA. CONTAK RENEWAL 4 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.

Recall number
Z-0950-05
Recalling firm
Guidant Corporation
Product
CONTAK RENEWAL 4 CRT-D (model no. H190 and H195) and CONTAK RENEWAL 4 HE CRT-D (model no. H197 and H199). Cardiac Resynchronization Therapy Defibrillator. Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 5511-5798 USA. CONTAK RENEWAL 4 devices provide ventricular tachyarrhythmia and cardiac resynchronization therapies.
Status
Terminated
Initiated
2005-06-24
Posted
2005-06-30
Terminated
2006-12-31
Reason
Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.
Root cause
Other
Action
Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.
Quantity
7,490
Distribution
worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Denmark, Eygpt, Finland, France, Germany, Greece, Guadeloupe, Guam, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Jordan, Kuwait, Labanon, Luxembourg, Marshall Islands, Mexico, Netherlands, New Zealand, Northern Mariana Islands, Norway, Peru, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom and Venezuela.
Lot, serial or model codes
all serial numbers
510(k) numbers
not on file
PMA numbers
not on file
Product code
NIK
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register