FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc: Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord, CA Medical charged-particle radiation therapy system.
- Recall number
- Z-0949-2009
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Product
- Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord, CA Medical charged-particle radiation therapy system.
- Status
- Terminated
- Initiated
- 2008-12-20
- Posted
- 2009-01-30
- Terminated
- 2010-12-09
- Reason
- The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.
- Root cause
- Device Design
- Action
- On 12/20/08, the firm issued letters via certified mail to direct consignees, informing them of the affected product and providing instructions on the recall. Contact Siemens Medical Solutions USA, Inc. at 1-925-246-8407 for additional information.
- Quantity
- 43 units
- Distribution
- Nationwide and Worldwide
- Lot, serial or model codes
- All units of Part Number: 8139789
- 510(k) numbers
- ["K072485"]
- PMA numbers
- not on file
- Product code
- IYE
- Firm FEI
- 2937457
- Firm city
- Concord
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as