Skip to the content

FDA Medical Device Recalls (openFDA)

Ecolab Inc: Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park, Illinois

Recall number
Z-0948-2014
Recalling firm
Ecolab Inc
Product
Vital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20, Manufactured for Vital Care Reps, Inc., Tinley Park, Illinois
Status
Terminated
Initiated
2013-07-02
Posted
2014-02-10
Terminated
2014-05-14
Reason
Certain lots of these Bur Hole Probe Covers did not undergo endotoxin (LAL) testing to the proper levels.
Root cause
Process design
Action
The recall was conducted to the consumer/user level. The firm notified their direct consignees of the problem and the recall by written letter mailed on July 2, 2013. The notice requested the consignees to quarantine any recalled product on hand, and complete and return the response form by July 31, 2013. The consignees who further distributed the recalled product were asked to notify their consignees.
Quantity
6,000 individual units
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Germany, Finland, Ireland, France, Spain, Great Britain, Belgium, Netherlands, Switzerland, Jordan, United Kingdom, Saudi Arabia, Hungary, Latvia, Malta, Norway, Poland, Portugal, and Canada.
Lot, serial or model codes
Lot Numbers: D111571, D120421, D123471, D83111, DA90721, and DA90961
510(k) numbers
["K882724"]
PMA numbers
not on file
Product code
KKX
Firm FEI
1000289767
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register