Philips Medical Systems (Cleveland) Inc: Brilliance Big Bore Model #728243 and #728244, Philips Medical Systems. Computed Tomography X-Ray System intended to produce of x-ray transmission data taken at different angles and planes.

Recall number
Z-0946-2012
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Brilliance Big Bore Model #728243 and #728244, Philips Medical Systems. Computed Tomography X-Ray System intended to produce of x-ray transmission data taken at different angles and planes.
Status
Terminated
Initiated
2011-12-13
Posted
2012-02-03
Terminated
2013-02-22
Reason
The Block and Multi Leaf Collimator (MLC) shape is not propagated to the beam that is modified by the Oppose operation. The contours can appear with the incorrect color in the Beam's Eye View Viewer, resulting in organs being incorrectly placed or incorrectly shaped.
Root cause
Software design
Action
On 12/13/2011, the firm sent Field Safety Notices, entitled URGENT - Medical Device Correction, to their customers. The letter identified the affected product, the problem, the hazard involved, and the actions to be taken by customers and the firm. Customers were to follow the instructions provided in order to address or avoid the issues. Philips will be installing a software update to address the identified issues free of charge. If customers need further information or support in regards to these issues, they are to contact their local Philips representative or local Philips Healthcare office. In North America and Canada, customers can contact the Customer Care Solutions Center at 1-800-722-9377.
Quantity
142 Units
Distribution
Worldwide Distribution -- US, including the states of AL, AZ, CA, CO, CT, FL, IA, IL, IN, KY, MA, MD, MI, MN, MO, NC, NH, NM, NY, OH, OR, PA, PR, TN, TX, VA, & WA and countries of AUSTRALIA, BELGIUM, BRAZIL, CANADA, CHINA, DENMARK, DOMINICAN REPUBLIC, FRANCE, GABON, GERMANY, HUNGARY, ISRAEL, ITALY, JORDAN, LUXEMBOURG, MALI, MEXICO, NETHERLANDS, OMAN, SLOVAKIA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, UKRAINE, UNITED ARAB EMIRATES, & UNITED KINGDOM.
Lot, serial or model codes
Model # 728243, Serial # 7006, 7009, 7060, 7068, 7081, 7095, 7122, 7127, 7133, 7150, 7154, 7185, 7233, 7245, 7278, 7285, 7298, 7337, 7395, 7400, 7450, 7465, 7468, 7506, 7518, 7520, 7521, 7523, 7524, 7528, 7545, 7547, 7560, 7568, 7569, 7573, 7578, 7581, 7584, 7591, 7593, 7597-7599, 7602, 7611, 7616, 7617, 7618, 7620, 7622-7635, 7637-7650, 7652, 7655-7659, 7661-7663, 7665-7667, 7671-7675, 7677-7682, 7684-7687, 7689-7691, 7693, 7695, 7696, 7698-7707, & 75000.  Model # 728244, Serial # 7543, 7571, 7600, 7619, 7621, 7636, 7651, 7653, 7654, 7660, 7668-7670, 7676, 7683, 7688, 7694, 7697, 7708, & 7710.
510(k) numbers
["K033357"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register