DSAART LLC: Brand Name: AART Gluteal Implant Product Name: Gluteal Implant Model/Catalog Numbers: 501-101 - Gluteal Implant style 1 size 1 UDI-DI code: B490501101 501-103 - Gluteal Implant style 1 size 3 UDI-DI code: B490501103 501-101 UDI-DI code: B490501101 501-103 - Gluteal Implant style 1 size 3 UDI-DI code: B490501103 501-105 - Gluteal Implant style 1 size 5 UDI-DI code: B490501105 501-206 - Gluteal Implant style 2 size 6 UDI-DI code: B490501206 501-301 - Gluteal Implant style 3 size 1 UDI-DI code: B490501301 501-302 - Gluteal Implant style 3 size 2 UDI-DI code: B490501302 501-303 - Gluteal Implant style 3 size 3 UDI-DI code: B490501303 501-304 - Gluteal Implant style 3 size 4 UDI-DI code: B490501304 501-305 - Gluteal Implant style 3 size 5 UDI-DI code: B490501305 501-306 - Gluteal Implant style 3 size 6 UDI-DI code: B490501306 501-307 - Gluteal Implant style 3 size 7 UDI-DI code: B490501307 501-309 - Gluteal Implant style 3 size 9 UDI-DI code: B490501309 501-310 - Gluteal Implant style 3 size 10 UDI-DI code: B490501310

Recall number
Z-0945-2026
Recalling firm
DSAART LLC
Product
Brand Name: AART Gluteal Implant Product Name: Gluteal Implant Model/Catalog Numbers: 501-101 - Gluteal Implant style 1 size 1 UDI-DI code: B490501101 501-103 - Gluteal Implant style 1 size 3 UDI-DI code: B490501103 501-101 UDI-DI code: B490501101 501-103 - Gluteal Implant style 1 size 3 UDI-DI code: B490501103 501-105 - Gluteal Implant style 1 size 5 UDI-DI code: B490501105 501-206 - Gluteal Implant style 2 size 6 UDI-DI code: B490501206 501-301 - Gluteal Implant style 3 size 1 UDI-DI code: B490501301 501-302 - Gluteal Implant style 3 size 2 UDI-DI code: B490501302 501-303 - Gluteal Implant style 3 size 3 UDI-DI code: B490501303 501-304 - Gluteal Implant style 3 size 4 UDI-DI code: B490501304 501-305 - Gluteal Implant style 3 size 5 UDI-DI code: B490501305 501-306 - Gluteal Implant style 3 size 6 UDI-DI code: B490501306 501-307 - Gluteal Implant style 3 size 7 UDI-DI code: B490501307 501-309 - Gluteal Implant style 3 size 9 UDI-DI code: B490501309 501-310 - Gluteal Implant style 3 size 10 UDI-DI code: B490501310
Status
Open, Classified
Initiated
2025-09-26
Posted
2025-12-18
Terminated
not on file
Reason
Due to manufactured products that deviated from established specifications and being distributed without obtaining the required FDA premarket approvals or clearances.
Root cause
No Marketing Application
Action
On 09/26/2025, the firm emailed a notice to an affected customer informing them that due to identified testing issues, the firm was requesting return of gluteal implants. On 11/17/2025, the firm send an "URGENT Medical Device removal of AART Silicone Implants" Letter to customer informing them that Alpha Aesthetics is voluntarily removing products as they conduct an analysis to determine the regulatory status of the devices. Customers are instructed to: 1. Either return the affected products that have not been implanted to Alpha Aesthetics at the company's expense, or discard affected products. 2. Acknowledge their actions via the enclosed "Medical Device Return Response" letter/form. For questions or further assistance with this removal, contact ** or email **.
Quantity
118 implants
Distribution
U.S.: AR, AZ, CA, CO, CT, DC, FL, GA, KY, IL, MA, MD, MI, MN, MS, NE, NJ, NY, OH, OK, PA, PR, TX, UT, VA, WA, and WI O.U.S.: N/A
Lot, serial or model codes
Gluteal Implant Model/Catalog Number: 501-101 UDI-DI code: B490501101 Lot Number: 24-01-030 Model/Catalog Number: 501-103 UDI-DI code: B490501103 Lot Numbers: 24-12-003 25-06-014 25-07-022 Model/Catalog Number: 501-105 UDI-DI code: B490501105 Lot Numbers: 24-07-012 24-09-013 Model/Catalog Number: 501-206 UDI-DI code: B490501206 Lot Number: 24-06-017 Model/Catalog Number: 501-301 UDI-DI code: B490501301 Lot Numbers: 23-10-020 25-01-021 Model/Catalog Number: 501-302 UDI-DI code: B490501302 Lot Numbers: 24-12-001 25-01-016 25-03-003 Model/Catalog Number: 501-303 UDI-DI code: B490501303 Lot Numbers: 24-10-004 24-11-001 25-01-017 25-02-011 25-03-003 25-04-009 Model/Catalog Number: 501-304 UDI-DI code: B490501304 Lot Numbers: 24-01-005 24-10-005 24-02-018 24-12-016 Model/Catalog Number: 501-305 UDI-DI code: B490501305 Lot Numbers: 24-09-009 25-01-001 25-03-004 25-04-010 Model/Catalog Number: 501-306 UDI-DI code: B490501306 Lot Numbers: 24-10-011 25-01-002 25-06-027 Model/Catalog Number: 501-307 UDI-DI code: B490501307 Lot Numbers: 24-09-010 24-12-002 25-04-007 Model/Catalog Number: 501-309 UDI-DI code: B490501309 Lot Number: 24-02-022 Model/Catalog Number: 501-310 UDI-DI code: B490501310 Lot Numbers: 24-10-010 25-02-009
510(k) numbers
["K021839"]
PMA numbers
not on file
Product code
MIB
Firm FEI
3003897287
Firm city
Carson City
Firm state
NV
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register