AGFA Corp.: IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000, and E2ZCK000. IMPAX Cardiovascular is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. The IMPAX Centric Polling Procedure provides a mechanism to ensure that patient demographic and merge/split operations performed on the IMPAX database utilizing the IMPAX Service Tool are captured and applied to the IMPAX CV database, in order to keep patient updates from the IMPAX database synchronized with the IMPAX CV database.
- Recall number
- Z-0945-2009
- Recalling firm
- AGFA Corp.
- Product
- IMPAX Cardiovascular Polling Procedure version 1.00.09, used in IMPAX 6.2.1 and 6.3.1.; Catalog numbers: E18R4000, L5NQJ000, L57OP000, LBCDB000, and E2ZCK000. IMPAX Cardiovascular is a picture archiving and communications system intended to be used as a networked cardiovascular information management system. The IMPAX Centric Polling Procedure provides a mechanism to ensure that patient demographic and merge/split operations performed on the IMPAX database utilizing the IMPAX Service Tool are captured and applied to the IMPAX CV database, in order to keep patient updates from the IMPAX database synchronized with the IMPAX CV database.
- Status
- Terminated
- Initiated
- 2008-12-24
- Posted
- 2009-02-03
- Terminated
- 2010-04-01
- Reason
- Polling procedure may fail when multiple merge/split is carried out.
- Root cause
- Software design
- Action
- All sites potentially affected by this product were notified of the issue directly via telephone. Each site was directed to observe a ten minute polling window as an interim mitigation until the normal mitigation by software Service Update is performed at the site by Agfa Support. An Urgent Safety Notice was sent via FED-EX on/about 12/24/2008 to all sites. The letter describes the potential issue and both interim and formal mitigation. Acknowledgment, via FAX-Back, that the information was received and understood was requested from the sites.
- Quantity
- 5 units
- Distribution
- Nationwide Distribution --- including states of FL, IL, OR, SC, and SD.
- Lot, serial or model codes
- Serial number: KQRCPF9, Equipment number: 15013696, ; Serial number: BQWBZ81, Equipment number: 10281800; Serial number: FB5C561, Equipment number: 10262174; Serial number: USE708NBP3, Equipment number: 15003839; and Serial number: GN6Z1E1, Equipment number: 15009724.
- 510(k) numbers
- ["K022292","K050228"]
- PMA numbers
- not on file
- Product code
- LLZ
- Firm FEI
- 3001236302
- Firm city
- Greenville
- Firm state
- SC
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28